Attaching the U-DPCAD
(Previous Page)
1. Using the Allen screwdriver provided with the microscope frame, loosen
the clamping screw on the trinocular observation tube or side port.
2. Fit the mount dovetail of the U-DPCAD and lightly tighten the clamping
screw for temporary clamping.
Rotate the U-DPCAD to collect inclination, then tighten the clamping
screw firmly.
Attaching the C-Mount Adapters
(Fig. 1)
1. Loosen the two clamping screws @ of the C-mount adapter for the
transmitted light port using the Allen screwdriver, then remove the
C-mount adapter ².
}The C-mount adapter for the transmitted light port has an identification
groove on the bottom side.
2. Screw the C-mount adapter firmly into the video camera to be used.
3. Place the C-mount adapter with video camera on the original position
and lightly tighten the clamping screws for temporary clamping. After
completing the parfocality adjustment, tighten the clamping screws firmly.
}Attach the C-mount adapter for the reflected light port in the same way
as above.