24
4. Place the revolving nosepiece mount ³ (with its objective mounts |
facing upward) where the objective mount has been by aligning the screw
holes.
Using the Allen wrench (for M3 screws), clamp the revolver mount with
the provided three clamping screws (M3, 6 mm long)
5
. (Fig. 28)
# As the screw holes may be hidden behind the objective mounts |,
clamp the nosepiece mount while rotating it. (Fig. 28)
5. Install the focusing assembly on the pillar, and install the observation
tube in the original position.
6. Attach two objectives onto their respective objective mounts | by screw-
ing. (Fig. 28)
Hold the objective and gently rotate it until a click position on the front,
where the objective to be used is engaged in the light path.
Operation for Optical Axis Aligned Photomicrography
}Align the center of objective and that of the illumination light path with
the photo light path (the right line) to enable photomicrography with high
contrast.
1. Rotate the desired objective @ clockwise (15°) till the click position for
the photo light path. (Fig. 30)
Fig. 28
³
|
5
6
Fig. 29
Fig. 30
@
3
Operation
(Fig. 29)