Olympus SZ2-STU1 Instructions Manual Download Page 1

INSTRUCTIONS

A X 6 6 2 3

This instruction manual is for the Olympus Stands for stereo microscopes. To ensure the safety,

obtain optimum performance and to familiarize yourself fully with the use of this equipment, we

recommend that you study this manual thoroughly before operating the system.

Retain this instruction manual in an easily accessible place near the work desk for future reference.

UNIVERSAL STAND TYPE 1

UNIVERSAL STAND TYPE 2

TABLE CLAMP STAND

SZ2-STU1

/

SZ2-STU2

SZ2-STU3

Summary of Contents for SZ2-STU1

Page 1: ...mance and to familiarize yourself fully with the use of this equipment we recommend that you study this manual thoroughly before operating the system Retain this instruction manual in an easily accessible place near the work desk for future reference UNIVERSAL STAND TYPE 1 UNIVERSAL STAND TYPE 2 TABLE CLAMP STAND SZ2 STU1 SZ2 STU2 SZ2 STU3 ...

Page 2: ......

Page 3: ... stand is provided with electroconductive coating in their external finish to reduce the surface resistance Connecting a grounding lead wire to the stand and also the horizontal arm in case of the SZ2 STU2 makes it possible eliminate static electricity quickly CAUTION To maintain the ESD protection performance always use the stand in combination with a stereo microscope arm used with microscope fo...

Page 4: ... hori zontal arm Figs 1 2 When the horizontal arm is extended at maximum the load to the bonder arm is limited as shown in the following table Front 70 Max 4 6 kg Side 85 Max 3 0 kg Rear 120 Max 2 0 kg The horizontal arm is most stable when it is oriented in the front 70 range shown in Fig 2 4 When a wide space is required in front of the microscope the horizontal arm should often be oriented in t...

Page 5: ...d capacity is 4 6 kg regardless of the orientation 3 Before installing or uninstalling the stand on or from the table remove the column from the table clamp to secure safety If you do not remove the column always have another person hold the horizontal arm section 4 If internal oscillations are troublesome insert the provided rubber sheet below the stand Fig 4 SZ2 STU3 ...

Page 6: ... section weight 2 kg Total weight 9 9 kg including the 0 9 kg bonder arm Mountable Modules To maintain the ESD protection performance the following modules should also be provided with the ESD protection performance Microscope body SZ2 series SZX7 SZ series or SD SF series Oblique illumination system LSGA mountable on the SZ2 STB1 Other modules mounted on the microscope body CAUTION From the viewp...

Page 7: ...ttach the column by firmly tightening the screws using the Allen wrench Insert the drop prevention ring below the horizontal arm in close contact with it It is convenient to store the Allen wrench in the Allen wrench storage position on the horizontal arm for future use 2 Attaching the Bonder Arm Fig 6 1 Lift the operation mistake prevention ring on the shaft at the distal end of the horizontal ar...

Page 8: ...rm section 3 To move the horizontal arm upward keep the drop prevention ring clamp ing knob tightened but loosen the vertical clamping knob After lifting the horizontal arm to the desired position tighten the vertical clamping knob again And then move up the drop prevention ring till immediately be neath the horizontal arm section and clamp the ring again To move the horizontal arm downward always...

Page 9: ...o slip out of the horizontal arm The stability is degraded if the distal end is attached on the side or above the horizontal arm In this case take care to increase the stability of the base plate Fig 10 Fig 11 6 To change the angle of the bonder arm loosen the bonder arm angle adjustment screw Fig 11 For the adjustment of the bonder arm angle read the instruction manual for the Arm Used with Micro...

Page 10: ... 5 8 kg The figures inside are the loads when the SZX stand arm weighing 1 2 kg is used Under this condition the stand tends to topple down easily if an other load is applied during the microscope operation It is rec ommended to set the horizontal arm as short as possible When the mounted microscope is heavy set the horizontal arm rotation stopper for the front 90 positioning For the method of thi...

Page 11: ...be provided with the ESD protection performance Microscope body SZX10 16 microscope focus adjustment module SZ2 STS SZ2 series SZX7 SZ series or SD SF series Oblique illumination system LSGA mountable on the SZ2 STS Other modules mounted on the microscope body CAUTION It is possible to mount the photography unit except for the large camera back or TV camera on the stand SZX Stand Arm SZ2 STS Unive...

Page 12: ...ct the ground lead wire Bare wire x 1 Drop prevention ring SZX extension adapter SZX10 16 body Microscope body drop prevention screw Column Base plate Allen wrench Clamping screws x 2 Connect the ground lead wire Fix the column securely by tightening both of the two clamping screws firmly using the provided Allen wrench To maintain the ESD protection performance be sure to ground the equipment usi...

Page 13: ...the arrow to clamp the extension adapter firmly 3 Mount the SZX microscope body and tighten the focusing assembly clamping knob 4 Screw in the microscope body drop prevention screw tightly using a coin 1 Remove the microscope body drop prevention screw loosen the clamp ing knob of the SZ2 STS attach the SZ2 STS to the mount shaft and tighten the clamping knob 2 Screw in the microscope body drop pr...

Page 14: ...of the drop prevention ring so that it points the front of the stand Fig 17 To ensure safety use the most stable rotation range unless other range is required for a special purpose Fig 16 Fig 17 Fig 18 3 To move the horizontal arm upward keep the drop prevention ring clamped but loosen the vertical clamping lever After lifting the horizontal arm to the desired position tighten the vertical clampin...

Page 15: ... lever revolve the mount shaft by 180 so that it points upward and mount the microscope body above the horizontal arm Fig 20 In this case the stability is deteriorated extremely Do not mount a heavy microscope body SZX in this way or mount a photographic unit on the microscope body mounted in this way 6 To rotate the microscope body horizontally loosen the clamping mechanism of the SZX stand arm o...

Page 16: ...cedure Take care so that the mount shaft is not tilted beyond the vertical positioning If the tilting clamping lever gets loose while the photography unit or TV camera is mounted it may topple down toward the observer To prevent this danger be sure to perform the following setup a Loosen the horizontal shaft rotation clamping lever withdraw the distal end of the arm by 2 to 3 mm and rotate the dis...