Before use, always inspect the power cable and plug for damage. The power
controller should not be used if visible damage is present. Use of damaged
components may be a safety hazard.
Only use the power controller with a properly grounded source. The safety of the
power controller relies on the provision of a proper ground connection. In
environments with moisture present, a ground fault circuit interrupter (GFCI) must
be used to ensure operator safety.
Some generators or DC-AC inverters may introduce significant levels of noise to the
system. This may degrade overall system performance or reduce the system life
expectancy. Use of generators or DC-AC inverters is not recommended and are used
at the operator’s risk.
The power connection [(3) in Figure 4-29 on page 113] of the power controller is used
to connect the power controller to a suitable 100–240 VAC, 50/60 Hz grounded power
source capable of supplying a minimum of 5 A.
4.5
Umbilical
FALLING OBJECT HAZARD. Ensure the umbilical can freely uncoil during
operation and does not become snagged. If the umbilical becomes snagged, the
SteerROVER may fall and SEVERE INJURY or DEATH could result.
CAUTION
NOTE
WARNING
Summary of Contents for SteerROVER
Page 8: ...10 004209 01EN Q7750159 Rev 2 May 2018 Table of Contents viii...
Page 10: ...10 004209 01EN Q7750159 Rev 2 May 2018 List of Abbreviations x...
Page 18: ...10 004209 01EN Q7750159 Rev 2 May 2018 Labels and Symbols 8...
Page 38: ...10 004209 01EN Q7750159 Rev 2 May 2018 Introduction 28...
Page 52: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 1 42...
Page 66: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 2 56...
Page 208: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 4 198...
Page 212: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 5 202...
Page 222: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 6 212...
Page 224: ...10 004209 01EN Q7750159 Rev 2 May 2018 Chapter 7 214...