Manual mode
n
Selecting the Manual button on the home screen allows the user to go
directly to the Treatment Screen to manually establish unique treatment
parameters, rather than selecting one of the preset treatment parameters.
n
Manual mode will default to the following parameters on both the right
and left pedals: Energy 0.2 J; Frequency 100 Hz; Pulse Width short.
n
Once adjusted, the settings can be saved.
View/export laser operation or laser therapy information
n
Press the Settings button on the Treatment Screen to enter the Settings
screen; then press Treatment Logs to enter the Treatments Logs screen.
n
Insert a USB memory stick into an
available USB port on the rear of the laser
system (circled below). Tap Export Logs
to save files to an external drive.
Other Important Features
n
The blast shield must be properly installed for the laser to operate. Each
laser system will come with a blast shield installed and two replacement
blast shields included in the box. Inspect the blast shield annually and
replace if necessary. Reference the User Manual for instructions on blast
shield replacement.
n
An operating door interlock connector is provided with each laser
system. This must be setup only by Olympus authorized technicians.