Olympus Single Use Cannula V StarTip V Manual Download Page 16

An Essential Tool to achieve a New Level of Efficiency 
in Biliary-Pancreatic Procedures.  Designed to Enhance 
the Revolutionary Exchange Capability of the V-System. 

130mm

Ring-marker section

Spiral-marker section

Ring-markers are visible from 50mm 
to 120mm from the distal end to 
help determine duct penetration.

50mm

120mm

400mm

2700mm / 4500mm

PTFE Coating

50mm

120mm

400mm

2700mm / 4500mm

PTFE Coating

Hydrophilic Coating Length:

500mm

Hydrophilic Coating Length:

500mm

All Designs are Optimized to Assure Smooth, 
Efficient Guidance of Devices to the Bile And Pancreati

Two-step  markers ensure visibility 
in the endoscopic field of view

Two-step markers on the Linear GuideV make 
it easy to confirm the movement and position 
of the guidewire within the endoscopic field of view.

The LinearGuideV has been designed to provide 
superior insertion capability, maximum endoscopic 
visibility, and easy device exchange.  When the 
LinearGuideV’s spiral markings are visible from 
the scope tip, the guidewire can be locked to 
simplify device exchange. 

Endoscopic visibility of the spiral-
markers indicates that 
LinearGuideV may be locked in 
the V-Groove.

The Innovative V-System Design Lets You Proceed 
with Confidence and Efficiency

Schematic of the V-Groove locking mechanism

V-System device replacement procedure 

Conventional scope (TJF-160)

V-System scope (TJF-160V)

Linear GuideV

Guide wire

V-Groove

The V-System is a complete system that integrates Olympus endoscopes and EndoTherapy devices. 
The revolutionary V-System design offers the option of guidewire manipulation by the physician 
or the assistant, allows easier exchange of catheters, and enhances cannulation capability.

Confirm the position of the V-Marking 

  on the V-System EndoTherapy accessory.  

When the V-Marking is completely visible 

  above the instrument channel port, lift the 

   forceps elevator to lock the guidewire.

The guidewire is now locked 

  into the V-Groove.

Completely remove the device.

The V-Groove in the V-Scope forceps 
elevator locks LinearGuideV in place 

When a guidewire slips out of position, it can 
be extremely frustrating. With the new 
Linear GuideV, unwanted movement is 
a thing of the past. Olympus’ V-System 
scopes feature a revolutionary V-Groove 
in the V-Scope foreceps elevator that 
allows Linear GuideV to be securely 
locked in place without any special 
attachment when extended 130mm from 
the distal end of the scope.  The approach 
to the bile or pancreatic ducts via the 
papilla can be accomplished quickly and easily 
without worrying about  the guidewire slipping.

Summary of Contents for Single Use Cannula V StarTip V

Page 1: ...Single Use Cannula V Single Use Cannula V...

Page 2: ...when to raise the elevator to lock the guidewire C Hook Now endoscopists have the option to manipulate guidewires and devices V Marking Indicates when to raise and lower the V Groove forceps elevator...

Page 3: ...Tower Gong Yuan No6 Royal Palace No6 Gon YuanXijie JianGuoMenNei DongCheng District Beijing 100005 China 117071 Moscow Malaya Kaluzhskaya 19 bld 1 fl 2 Russia 31 Gilby Road Mount Waverley VIC 3149 Au...

Page 4: ...Single Use 3 Lumen Extraction Balloon V...

Page 5: ...that the device tip has reached the distal end of the scope and the V Groove forceps elevator may be lowered When withdrawing the device from the scope the same marking indicates when to raise the el...

Page 6: ...yringes for reliable inflation Multiple balloon sizing Injection Lumen 0 035 Guidewire Lumen Air Lumen For easy confirmation of balloon position during fluoroscopy two radiopaque bands are incorporate...

Page 7: ...ions design and accessories are subject to change without any notice or obligation on the part of the manufacturer Shinjuku Monolith 3 1 Nishi Shinjuku 2 chome Shinjuku ku Tokyo 163 0914 Japan Postfac...

Page 8: ...Single Use Retrieval Basket V Single Use Retrieval Basket V...

Page 9: ...ng is located on the proximal side of the sheath When this marking reaches the channel port on the scope s control section it indicates that the device tip has reached the distal end of the scope and...

Page 10: ...ngapore 248373 Room 1520 1527 Ocean Centre 5 Canton Road Tsimshatsui Kowloon Hong Kong Room 1401 East Tower Gong Yuan No6 Royal Palace No6 Gon YuanXijie JianGuoMenNei DongCheng District Beijing 100005...

Page 11: ...Single Use 2 Lumen Sphincterotome V Single Use 3 Lumen Sphincterotome V...

Page 12: ...ed on the proximal side of the sheath When this marking reaches the channel port on the scope s control section it indicates that the device tip has reached the distal end of the scope and the V Groov...

Page 13: ...ases in which cannulation is difficult due to strictures or when insertion into the minor papilla is required The tapered tip CleverCut3V is compatible with a 0 025 diameter guidewire The distal ends...

Page 14: ...t Beijing 100005 China 117071 Moscow Malaya Kaluzhskaya 19 bld 1 fl 2 Russia 31 Gilby Road Mount Waverley VIC 3149 Australia 2 Corporate Center Drive Melville N Y 11747 3157 U S A Spacifications Model...

Page 15: ...DISPOSABLE GUIDEWIRE...

Page 16: ...sign Lets You Proceed with Confidence and Efficiency Schematic of the V Groove locking mechanism V System device replacement procedure Conventional scope TJF 160 V System scope TJF 160V LinearGuideV G...

Page 17: ...5 2 and 5 2 5 2 when used with an Olympus sphinctetorome Removal of the guidewire is not necessary during sphincterotomy All Designs are Optimized to Assure Smooth Efficient Guidance of Devices to th...

Page 18: ...oint Office Tower Singapore 248373 Room 1520 1527 Ocean Centre 5 Canton Road Tsimshatsui Kowloon Hong Kong Room 1401 East Tower Gong Yuan No6 Royal Palace No6 Gon YuanXijie JianGuoMenNei DongCheng Dis...

Reviews: