Olympus R1 Quick Reference Manual Download Page 2

 

 
 

This Quick Reference Guide provides instructions to use the Cell Counter model R1. For detailed instructions, please refer to the user manual in the supplied USB memory. 
  

Intended Use

 

The R1 is an electrical laboratory instrument for scientific research use only. 
It is not a medical, therapeutic, or in vitro diagnostics device. 
 

General Guidelines

 

Follow the instructions below to obtain the best results with the R1. 

1. 

The R1 accurately detects and counts cells 3-60 µm in size at concentrations ranging from 
5 x 10

4

 to 1 x 10

7

 cells/mL. 

 

2. 

Perform cell counting within three minutes of mixing samples with trypan blue for accurate 
cell viability measurements. If necessary, count your sample twice (duplicate readings) and 
take an average.  
 

3. 

Hold Cell  Counting Slides by the edges to avoid touching the optical surface. Ensure that 
the optical surfaces of the slide do not become smudged, damaged, or contaminated. 
 

4. 

Do not reuse Cell Counting Slides. Used slides must be disposed as biohazardous waste 
according to the rules and regulations of your local government.   

5. 

As the R1 is calibrated before shipping, recalibration before use is not necessary. However, 
if recalibration is needed, please refer to Section 2.3.3: Settings: Background Calibration in 
the user manual.  

 

Safety Precautions 

Before using this instrument, read the user manual carefully to ensure that you know how to operate 
it safely and correctly. To protect yourself and others from personal injury or damage to property, it is 
essential that you read the warnings and information provided. Use the instrument as specified by 
Olympus.  

1. 

Install the instrument on a sturdy and level surface. Avoid vibrations from other devices. 
 

2. 

Do not touch any of the components with wet hands. 
 

3. 

Use the components provided or authorized by Olympus. If the proper combination of 
components are not used, product safety performance cannot be guaranteed.  

Ferrite Core Placement

 

Take a ferrite core and open it. Place the plug side of the 
cable of the AC adapter in the ferrite core as shown in the 
figure. Close the ferrite core. 

 

 

The AC adapter cord and power cord are vulnerable to bending and twisting. Do not 
apply excessive force.   

 

Always use the AC adapter and power cord provided by Olympus. If the proper AC 
adapter and power cord are not used, the electric safety and EMC (Electro-Magnetic 
Compatibility) performance of the device cannot be guaranteed. 

 

Connect the ground terminal of the power cord and that of the power outlet. If the 
device is not grounded, our intended electric safety and EMC performance of the 
device cannot be guaranteed. 

 

www.olympus-global.com 

 

 

OLYMPUS CORPORATION 

Shinjuku Monolith, 2-3-1 Nishi-Shinjuku, Shinjuku-ku, Tokyo 163-0914, Japan 

Distributed by 

OLYMPUS EUROPA SE & CO. KG 

Wendenstrasse 14-18, 20097 Hamburg, Germany 

OLYMPUS SCIENTIFIC SOLUTIONS AMERICAS CORP. 

48 Woerd Avenue Waltham, MA 02453, U.S.A. 

OLYMPUS SINGAPORE PTE LTD. 

491B River Valley Road, #12-01/04 Valley Point Office Tower, Singapore 248373 

OLYMPUS AUSTRALIA PTY. LTD. 

3 Acacia Place, Notting Hill VIC 3168, Australia 

OLYMPUS KOREA CO., LTD. 

8F Olympus Tower, 446 Bongeunsa-ro, Gangnam-gu, Seoul, Korea 135-509 

 

This device complies with the requirements of both directive 2004/108/EC concerning 
electromagnetic compatibility and directive 2006/95/EC concerning low voltage. The CE 
marking indicates compliance with the above directives. This system is applied with the 
requirements of standard IEC/EN61326-1 concerning electromagnetic compatibility. 
Immunity: Applied to industrial environment requirements. 
Emission: Class A, applied to industrial environment requirements. 
Some interference may occur if this system is used in a domestic location.

 