background image

INSTRUCTIONS

ENDOSCOPIC FLUSHING PUMP

OFP

CAUTION

Federal (USA) law restricts this device to sale

by or on the order of a physician.

0086

Summary of Contents for OFP

Page 1: ...INSTRUCTIONS ENDOSCOPIC FLUSHING PUMP OFP CAUTION Federal USA law restricts this device to sale by or on the order of a physician 0086...

Page 2: ...taltic flushing pump intended for use in medical facilities under the direction of a trained physician The OFP has been designed to be used with Olympus EVIS EXERA gastrointestinal videoscopes having...

Page 3: ...G PRECAUTIONS 3 3 INSTRUCTIONS FOR USE 4 4 STEAM STERILIZATION AUTOCLAVING 9 5 CARE AND STORAGE 10 6 MAINTENANCE AND REPAIR 10 7 SPARES ACCESSORIES 12 8 SPECIFICATIONS 12 9 TECHNICAL DESCRIPTION 13 10...

Page 4: ...he standard set are present Contact your Olympus service centre or nearest Olympus office if any parts are damaged or missing FIGURE 1 1 3 1 4 2 5 Contents Qty 1 Olympus Flushing Pump OFP 1 2 Fluid co...

Page 5: ...ote control cable connection Label 3 IEC power receptacle with integral fuses Potential equalisation terminal Fluid container strap Fluid container holder 1 3 Labels Label 1 Refer to Instructions Labe...

Page 6: ...e air will be supplied to the patient 2 12 To prevent free flow of fluid to the patient do not open the pump cover when the OFP is connected to an endoscope which is inserted into the patient and ensu...

Page 7: ...tion of the patient and use clinical judgement to set the flow rate from the pump to a suitable level to avoid patient trauma The flow control should always be set to MIN at the start of the procedure...

Page 8: ...rformance and will recover if left to rest 3 2 Place the OFP on a level flat surface below patient level ensuring the fluid container is not exposed to heat emitted from adjacent equipment such as a l...

Page 9: ...d auxiliary water inlet on the endoscope CAUTION Use only in conjunction with Olympus EVIS EXERA series endoscopes having an auxiliary water inlet and MAJ 855 auxiliary water tube Ensure the MAJ 855 h...

Page 10: ...delivered when using a 50ml syringe CAUTION NEVER allow the fluid container to run dry otherwise air will be supplied to the patient 3 12 To refill the fluid container during use leave the pump cover...

Page 11: ...isconnect the endoscope luer connector Figure 3 5b iv allow any fluid in the tube to return to the fluid container then remove the tube from the container v discard the irrigation tubing and filter at...

Page 12: ...ry Handling a wet package can compromise its sterility For details on operation of the autoclave refer to the instruction manual for the autoclave Typical parameters for pre vacuum autoclaves for wrap...

Page 13: ...ead Empty the fluid container and autoclave prior to storage 2 The OFP should be stored within the environmental conditions given in Section 8 6 MAINTENANCE AND REPAIR 6 1 Routine Maintenance The foll...

Page 14: ...eck the condition of both fuses located at the rear of the unit and the fuse in the power cable plug as appropriate CAUTION The fuse must always be replaced with one of the correct rating If the fuse...

Page 15: ...ty This product complies with the requirements of EN IEC 60601 1 2 for emissions and immunity and as such its operation is unlikely to be affected by or cause interference with equipment meeting appro...

Page 16: ...staltic Pump operation Water feed controlled by pneumatic footswitch Flow rate variable controlled from front panel Flow rate delivery from the pump will depend on the length and type of Olympus EVIS...

Page 17: ...d Check tubing and replace if necessary with specified tube see Section 7 Pump head or tube rollers damaged Switch unit OFF open pump cover and check for signs of damage If damage is evident refer to...

Page 18: ......

Page 19: ......

Page 20: ...01702 465677 Telephone 01702 616333 2951 Ishikawa cho Hachioji shi Tokyo 192 8507 Japan Fax 0426 46 2429 Telephone 0426 42 2111 OLYMPUS MEDICAL SYSTEMS CORP OLYMPUS SINGAPORE PTE LTD OLYMPUS MOSCOW LI...

Reviews: