The NORTEC 600 menus described in this section are opened by pressing the
corresponding menu keys, which are described in “Menu Keys” on page 77.
4.3.1
Frequency (FREQ 1) Menu — MAIN FILTER Key
FREQ (frequency)
The
FREQ
setting determines the frequency of the drive signal of the eddy
current probe. This is adjustable from 10 Hz (0.01 kHz) to 12 MHz.
To adjust the frequency setting, press the MAIN FILTER menu key (
),
followed by the A key. With
FREQ
highlighted, rotate the knob until the desired
frequency is displayed.
To speed up the frequency selection process, press the Enter key (
) once while the
FREQ
function is highlighted to enable the coarse knob function.
FREQ
will be
underlined when this function is enabled. To turn off the coarse-knob function, press
one more time.
ANGLE (rotation)
The phase angle (or rotation) of the eddy current signal is set using the ANGLE
key (
). By default, the angle is set in 1 degree increments, from 0 to
359 degrees.
To adjust the angle setting, press the MAIN FILTER menu key (
), followed by
the B key. With
ANGLE
highlighted, rotate the knob until the desired angle is
displayed.
TIP
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...