DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
211
Figure 5
‑
127 Position 3 on the standard
22. Move the probe over notch B, and then, while holding the probe at this position,
adjust the
V GAIN 2
(E key) until the mix signal extends upward from the null
point 4 vertical divisions (see Figure 5-128 on page 211).
Figure 5
‑
128 Adjusting the V GAIN 2 value
23. Press the MAIN FILTER menu key (
) repeatedly until the
FREQ 1
page is
displayed.
24. Place the probe on position 1, press the A-LIFT NULL key (
), and then move
the probe back and forth over the next fastener on the left.
3
B
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...