DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
199
Figure 5
‑
110 Adjusting the H MIX GN value
21. Increase the
V MIX GN
(D key) until the signal’s maximum amplitude reaches
80 % of the screen height (see Figure 5-111 on page 199).
Figure 5
‑
111 Adjusting the V MIX GN value
22. Press the FREEZE key (
) to unfreeze the acquisition.
23. Press the FULL NEXT key (
) to toggle to the full-screen mode.
24. Null the probe with the A-LIFT NULL key (
) on position 1, and then scan the
three corrosion defects.