DMTA-10040-01EN, Rev. E, February 2018
Chapter 5
150
Figure 5
‑
41 The Surface Cracks application
3.
Press the MAIN FILTER menu key (
) once, and then set the
FREQ
(A key) to
10 kHz using the knob.
4.
Press the ALARM menu key (
), and then deactivate
ALARM 1
(A key).
5.
Press the DISP menu key (
) twice, and then set the
GRID
(D key) to
10 × 10
using the knob.
6.
Press the DISP menu key (
), and then set the
H POS
(D key) to
100 %
using
the knob.
7.
Set the
V POS
(E key) to
20 %
using the knob.
To calibrate the signals
1.
Place the probe on the standard, between two notches (without any shim), and
then press and hold the A-LIFT NULL key (
) to activate the automatic lift-off
function.
After a brief moment, the NORTEC 600 beeps and displays
LIFT PROBE
at the
top of the screen. When this text is displayed, lift the probe into the air, and wait
for the message to disappear.
2.
Repeat step 1 until you become familiar with the automatic lift-off function.
It may take some practice to obtain the proper timing with the automatic lift-off
function, but once mastered, this function will help you obtain a faster calibration.
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...