background image

ISSUE6

7-1

Troubleshooting

MU-1

Chapter 7: Troubleshooting

Refer to the “Troubleshooting” of Instruction Manual.

Summary of Contents for MU-1

Page 1: ...ISSUE6 2016 04 149P 0015 MU 1 MAINTENANCE UNIT Maintenance Manual...

Page 2: ......

Page 3: ...dified product The information contained in this manual is subject to change at any time without notice Later editions of this manual will contain revised and updated material Applicable Unit MAINTENA...

Page 4: ......

Page 5: ...ous Components 1 5 6 System Chart 1 7 Chapter 2 Specifications 2 1 Chapter 3 Installation and Connection 3 1 Chapter 4 Care Storage and Disposal 4 1 Chapter 5 Inspection 5 1 1 Inspection Check Sheet 5...

Page 6: ......

Page 7: ...r an air bubbles 3 This product can be combined with the AW channel cleaning adapter and chemical tank to feed rubbing alcohol into the OES scope and EVIS scope air and water channel in order to dry t...

Page 8: ...o drain the water feeding channel cover the water supply connector with a finger Same as with the former model 4 The ambient operation conditions are as follows a Temperature 10 to 40 C b Humidity 30...

Page 9: ...temperature of 20 to 25 C 4 1 4 Control Method By turning the power switch ON OFF 4 1 5 Hook Installation Can be hung down by connecting the hanger to the hook installed in a wall The hook is screwed...

Page 10: ...Product Outline 1 4 ISSUE6 MU 1 4 2 3 Atmospheric Pressure 700 to 1060 Pa 4 2 4 Inflammable Atmosphere Use in an inflammable atmosphere is prohibited...

Page 11: ...air is supplied through the connecter Power supply switch Pushing the top will turn on the power supply and the power supply switch lamp will light at the same time The pump will start to operate and...

Page 12: ...Product Outline 1 6 ISSUE6 MU 1 5 2 Back Power supply inlet Breaker...

Page 13: ...1 6 System Chart Water resistant Cap MaintenanceUnit MU 1 Alcohol Container MB 264 only EVIS Videoscope 100 200 series OES Fiberscope 10 20 30 series Leakage Tester MB 155 OES Fiberscope EVIS Videosco...

Page 14: ......

Page 15: ...ISSUE6 2 1 Specifications MU 1 Chapter 2 Specifications Refer to the Specifications of Instruction Manual...

Page 16: ......

Page 17: ...ISSUE6 3 1 Installation and Connection MU 1 Chapter 3 Installation and Connection Refer to the Installation and Connection of Instruction Manual...

Page 18: ......

Page 19: ...ISSUE6 4 1 Care Storage and Disposal MU 1 Chapter 4 Care Storage and Disposal Refer to the Care Storage and Disposal of Instruction Manual...

Page 20: ......

Page 21: ...eck item may change with the destination and specifications There may be a check item which is on this list and is not in the instruction manual Please refer to the instruction manual Chapter of Inspe...

Page 22: ...MU 1 MU 1 INSPECTION CHECK SHEET Serial number Year Month Day 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Inspecting the Operation Inspector Day 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Inspecting the...

Page 23: ...esting instrument for the measurement methods 2 Grounding Leak Current Criteria Grounding leak currents should not exceed the values indicated below Inspection procedure Measure the leak current of th...

Page 24: ...and grounding line should be 0 2 ohm or less Inspection procedure Measure the resistance between the grounding terminal in the removable power supply cord and the screw that is farthest from the inlet...

Page 25: ...pected by Serial number Checked by Item Criteria Result 2 1 1 External leak current Normal Condition 100 A or less Single Fault 500 A or less 2 1 2 Grounding leak current Normal Condition 500 A or les...

Page 26: ......

Page 27: ...ISSUE6 7 1 Troubleshooting MU 1 Chapter 7 Troubleshooting Refer to the Troubleshooting of Instruction Manual...

Page 28: ...2951 Ishikawa machi Hachioji shi Tokyo Japan Medical Service Engineering Department Medical Repair Service Division PRINTED IN JAPAN FIRST EDITION 10 2003...

Reviews: