Problem
Cause
Remedy
Page
A light is not emitted.
The AC adapter is not connected.
Connect the outlet plug to the AC adapter
properly.
18
Make sure that the outlet plug is
connected to the commercial power
supply and the DC power jack is
connected to the LG-LSLED properly.
18
The power is not supplied to the AC
adapter.
Check the commercial power supply.
–
The power is not ON.
Turn ON the main switch.
10
The light guide is disconnected.
Connect the light guide to the LG-LSLED
(LED light source for light guide) and the
microscope properly.
15
The distal end of the light guide is
covered by the cap.
Remove the cap.
–
The light intensity is
reduced.
The light source unit is overheated.
Make sure that the cooling fan is working
properly.
–
Make sure that air vents or exhaust ports
are not covered by obstacles.
–
The light guide is about to be
disconnected.
Connect the light guide to the LG-LSLED
(LED light source for light guide) and the
microscope properly.
15
The filter is engaged in the light path.
Pull out the filter slider to the click position.
10
The filter slider is not inserted
completely to the end.
Make sure that the filter slider is inserted
completely to the end.
10
The distal end of light guide or the top
surface of filter is dirty.
Remove the dust/dirt.
7
4
Troubleshooting
If problems occur, please review the following list and take remedial action as needed.
If you cannot solve the problem after checking the entire list, please contact Olympus for assistance.
Repair request
If you cannot solve the problems even though taking actions described in Troubleshooting, please contact Olympus
for assistance. At that time, please tell them the following information as well.
· Product name and abbreviation (Example: LED light source for light guide LG-LSLED)
· Product number
· Problem
This section describes the possible problems of this product and their remedies.