H Y S T E R O F L O W   Q U I C K   R E F E R E N C E   G U I D E

5

C O N N E C T   O U T F L O W   T U B I N G

I N S T R U M E N T   R E C O G N I T I O N   ( I R )

1.

Open the inflow stopcock on the
 instrument.

2.

Close the outflow stopcock on the
 instrument.

3.

Hold the instrument at operative height.

4.

When prompted, press 

Yes

to begin

IR activation.

5.

Turn the inflow stopcock off when
IR is complete. Screen prompt and
 single beep indicate IR completion.

Do Not Press Stop!

1.

Place containers on the bottom scale
and toggle them with the connection
tubes.

2.

Connect the yellow Leur Lok on the
outflow tubing to the outflow port of
the instrument.

3.

Connect the short end of the Y tube
to the patient drape.

4.

Connect the long end of the Y tube
to the “Patient” port of the first
 container.

5.

Connect the last container tubing
to suction.

To Suction

Outflow

Summary of Contents for HysteroFlow/HysteroBalance II

Page 1: ...HYSTEROFLOW HYSTEROBALANCE II HYSTEROSCOPIC FLUID MANAGEMENT SYSTEM QUICK REFERENCE GUIDE Option 1 Technical Assistance Option 2 Customer Service Option 3 Repair Option 4 Field Inventory...

Page 2: ...ARNING This quick reference guide is only a summary of the HysteroFlow HysteroBalance II Instructions for Use For detailed operating instructions be sure to follow the Instruction Manual that was incl...

Page 3: ...ooks 4 Close inflow and outflow stopcocks on the instrument 5 Connect blue Leur Lok on the inflow tubing to the inflow port on the instrument 6 Press OK next to Operative Mode and press Start 7 Carefu...

Page 4: ...al flow value between 30 and 500 ml min Maximum Deficit Menu Select a deficit threshold value between 200 and 2500 ml Alarm Volume Menu Select a volume between 1 and 5 Bag Size Select a bag size betwe...

Page 5: ...ow stopcock off when IR is complete Screen prompt and single beep indicate IR completion Do Not Press Stop 1 Place containers on the bottom scale and toggle them with the connection tubes 2 Connect th...

Page 6: ...ressure Deficit Exceeded 10 Beeps In increments of 100 ml above the preset value Critical Overpressure 10 Beeps Actual Pressure 200 mmHg above nominal pressure Rapid Deficit 10 Beeps Deficit change 30...

Page 7: ...Outflow tubing set Sterile 5 pkg WA40635A Inflow tubing set Sterile 10 pkg 006775 901 Under Buttocks Drape Fluid Collection Set 5 pkg 006772 901 3L Canisters 36 pkg WA40634A Inflow and Outflow tubing...

Page 8: ...ot encourage or recommend any actions that deviate from or are not covered by the IFU 2013 Olympus America Inc All rights reserved Trademark and Registered Trademark of Olympus or its affiliated entit...