DMTA034-01EN, Rev. B, March 2014
Chapter 2
28
Figure 2-9 Adjusting the wedge’s carbide wear pins
To adjust the carbide wear pins on a wedge
1.
Turn off the couplant flow.
2.
Place the scanner upside down on a flat surface.
3.
Use a hexagonal key to adjust the height of the carbide wear pins on each
installed wedge. Ensure that the top of each pin is flush to the surface of the
wedge (see Figure 2-9 on page 28).
This concludes the procedure to adjust the carbide wear pins on a wedge.
2.3.2
Replacing a Conventional Ultrasonic Transducer
To replace a conventional ultrasonic transducer, the assembly holding the transducer
must be removed and partially disassembled.
To replace a conventional ultrasonic transducer
1.
Turn off the couplant flow.
2.
Disconnect the couplant tubes from the couplant source, or from the wedge.
Carbide wear pins
The carbide wear pins must be flush with
the surface of the wedge.
Bottom view
Top view
Use a hexagonal key to adjust
the height of the carbide
wear pins.