10-003591-01EN [Q7780031], Rev. 4, January 2020
Maintenance and Troubleshooting
51
To change the foam gasket
1.
Pull off the old gasket, and, if necessary, carefully clean the mounting surface (see
Figure 5-3 on page 51).
2.
Choose the gasket model that is compatible with the wedge diameter.
3.
Peel the backing off the adhesive on the new gasket.
4.
Align the adhesive side of the gasket with the mounting surface, and then press it
against the surface to install it.
Figure 5-3 Changing the foam gasket
To change the o-ring seal between the probe and wedge
1.
Use an appropriate tool such as a small flathead screwdriver to carefully lift and
pull off the old o-ring seal without damaging the mounting surface on the wedge
(see Figure 5-4 on page 52).
2.
Install the new o-ring seal.
Clean mounting surface
after removing old gasket.
Peel off the backing
on new gasket.
Align and press new gasket onto
mounting surface.
Summary of Contents for FlexoFORM
Page 6: ...Table of Contents vi 10 003591 01EN Q7780031 Rev 4 January 2020 ...
Page 8: ...10 003591 01EN Q7780031 Rev 4 January 2020 List of Abbreviations viii ...
Page 12: ...10 003591 01EN Q7780031 Rev 4 January 2020 Labels and Symbols 4 ...
Page 24: ...10 003591 01EN Q7780031 Rev 4 January 2020 Important Information Please Read Before Use 16 ...
Page 26: ...10 003591 01EN Q7780031 Rev 4 January 2020 Introduction 18 ...
Page 30: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 1 22 ...
Page 44: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 3 36 ...
Page 82: ...10 003591 01EN Q7780031 Rev 4 January 2020 Chapter 7 74 ...
Page 86: ...10 003591 01EN Q7780031 Rev 4 January 2020 Appendix B 78 ...
Page 90: ...10 003591 01EN Q7780031 Rev 4 January 2020 List of Tables 82 ...
Page 94: ...10 003591 01EN Q7780031 Rev 4 January 2020 Index 86 ...