Chapter 3 Cleaning, Disinfection and Sterilization Procedures
33
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3.5
Manual cleaning
To prevent a water leak, do not apply excessive force when
washing the endoscope.
After completing the leakage test, perform manual cleaning according to the
procedures described below.
In case of excessive bleeding and/or delayed reprocessing, first follow the
procedure described in “Presoak for excessive bleeding and/or delayed
reprocessing after each procedure” on page 47, then follow the procedure
described below.
Reusable parts which are normally reprocessed with the
endoscope
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
• Channel plug
(MH-944)
• Injection tube
(MH-946)
• Water-resistant cap
(MH-553)
• Chain for water-resistant cap
(MAJ-1119)
• Washing tube
(MH-974)
• Personal protective equipment
• Soft brush
• Clean lint-free cloths
• Large basin
• Detergent solution
• Clean water
• Suction cleaning adapter
(MH-856)
• Channel plug
(MH-944)
• Injection tube
(MH-946)
• Channel cleaning brush
(BW-20T)
• Channel-opening cleaning brush
(MH-507)
• Suction pump
• 30 cm
3
(30 ml) syringe
• Washing tube
(MH-974)
• 5 cm
3
(5 ml) syringe