Chapter 2 Instrument Nomenclature and Specifications
15
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
Specifications
Endoscope functions
Model
TJF-160VR
Optical system
Field of view
100
°
Direction of view
Backward
Sideviewing 5
°
Depth of field
5 – 60 mm
Insertion tube
Distal end outer
diameter
ø 13.5 mm
Distal end enlarged
1. Air/water nozzle
2. Objective lens
3. Light guide lens
4. Instrument channel outlet
5. Forceps elevator
6. Guidewire-locking groove
7. Elevator wire
8. Hook
9. White ring
Insertion tube outer
diameter
ø 11.3 mm
Working length
1240 mm
Instrument channel
Channel inner
diameter
ø 4.2 mm
Minimum visible
distance
10 mm
Direction from which
endo-therapy
accessories enter
and exit the
endoscopic image
UP
LEFT
RIGHT
DOWN
8.
3.
2.
1.
9.
4.
7.
5.
6.