Chapter 2 Instrument Nomenclature and Specifications
33
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
• The chain for water-resistant cap and water-resistant cap
cannot be steam sterilized (autoclaved); doing so can
damage them severely.
Figure 2.1
To ensure that you do not forget to attach the water-resistant
cap, it is recommended that you connect it to the
endoscope’s S-cord connector mount or the chain connector
of the GIF-N180 using the chain for water-resistant cap.
1.
Confirm that the chain for water-resistant cap is free from cracks, flaws,
wear, deformation or other damages (see Figure 2.1).
2.
Align the notch on the connecting plate with the pin on the venting connector
of the water-resistant cap (MH-553, see Figure 2.2).
3.
Place the connecting plate over the venting connector (see Figure 2.2).
4.
Confirm that the connecting plate is securely attached to the foot of the
venting connector and can be smoothly rotated.
5.
Place the hole on the fitting over the endoscope’s S-cord connector mount
or the chain connector of the GIF-N180 (see Figure 2.3).
Notch
Connecting plate
Chain part
Fitting part
Hole