Olympus ESG-300 Quick Reference Manual Download Page 2

CONTINUED ON NEXT PAGE

ESG-300

 (

Electrosurgical Generator)

 

Quick Reference Guide

Bipolar Home Screen

Active Mode Name

Power adjustment

Effect level adjustment

Open user defined settings

System setup

Output Socket Name

Change Output Sockets

Argon Home Screen

Foot Switch Indicator

Contact Quality Monitor

Output Socket Name

Argon Gas Flow Rate

Type of Connected Argon Probe

Argon Supply Indicator

Communication Indicator
Setting Name

Power Level

Effect Level

Summary of Contents for ESG-300

Page 1: ...ase Electrosurgical Generator ESG 300 Bipolar Monopolar On Off Neutral Argon Plasma Unit APU 300 On Off Argon Graphical User Interface Purge button Monopolar Home Screen Active socket Contact Quality...

Page 2: ...ment Effect level adjustment Open user defined settings System setup Output Socket Name Change Output Sockets Argon Home Screen Foot Switch Indicator Contact Quality Monitor Output Socket Name Argon G...

Page 3: ...eps 1 2a Turn on the ESG 300 APU 300 Select Mode Monopolar Bipolar Argon Monopolar Bipolar Argon Select mode by tapping on the desired mode located on the bottom of the touch screen 2b Optional Select...

Page 4: ...bar on the touch screen 2 In the menu screen tap to select the desired mode 3 Once selected the home screen will display the change 4 Changing the Power Level Select to adjust power level 5 Changing...

Page 5: ...the neutral electrode to the patient 2 Connect to the ESG 300 8 Monitor Neutral Electrodes Contact Quality Monitor If the red neutral electrode lights up alarm tone sounds Is the neutral electrode co...

Page 6: ...ESG 300 System Operating Steps 9 Bipolar Devices Monopolar Devices Neutral Electrodes Argon Plasma Probes Connect device in appropriate receptacle Creating User Defined Settings STEP 1 On the home sc...

Page 7: ...Screen OK STEP 1 On the Home screen Select Menu STEP 2 Edit Settings Procedures screen Select New Procedure STEP 3 New Procedures screen Use up down to highlight the setting Use arrow to move the sett...

Page 8: ...ForcedCoag 20 2 SoftCoag 50 3 Pre Cut Incision Knife PulseCut Fast 120 2 BlendCut 30 2 Submucosal Dissection Knife PowerCoag 30 2 Hemostasis Closed Knife PowerCoag 30 2 Peroral Endoscopic Myotomy POEM...

Page 9: ...ForcedArgon 50 N A Barrett s Esophagus Activate MAPC probe within a 2 5 mm distance from tissue PulsedArgon Fast 40 N A Pulmonary Procedures Procedure Instrument or Technique Mode Power Effect Less e...

Page 10: ...on Cutting of varying tissue structures with increased hemostasis capacities Superficial coagulation with increased dissection capability Superficial pinpoint coagulation High peak voltage for superfi...