DMTA-10055-01EN, Rev. A, February 2015
Chapter 7
124
Both gates can be used to take thickness measurements with straight-beam
transducers, sound path and depth measurements with angle beam transducers,
measure signal amplitude, measure time-of-flight in microseconds, or to trigger
threshold and minimum-depth alarms. The gates can also be used together to take
echo-to-echo thickness measurements.
You can control the gates using the parameters in the
Gate 1
and
Gate 2
menus (see
Figure 7-2 The Gate 1 menu
The available gate parameters are:
Zoom
Used to zoom the display on the width of the gate (see “Zooming on a Gate” on
page 132 for details).
Start
Used to adjust the gate start position.
Width
Used to adjust the gate width.
Level
Used to adjust the vertical position of the gate.