DMTA
‐
10007
‐
01EN
[U8778365],
Rev. B,
June
2010
Basic
Operation
41
7.
Use
the
knob
or
arrow
keys
to
adjust
the
displayed
value
to
the
appropriate
thickness.
For
this
example,
the
transducer
is
coupled
to
a
5 mm
step.
8.
Press
Continue
to
accept
the
adjusted
value,
and
then
move
on
to
the
second
calibration
step.
9.
Couple
the
transducer
to
the
thick
step
of
material.
10. Position
gate 1
around
the
resulting
indication
on
screen.
11. Bring
the
indication
to
80 %
full
screen
height
using
the
(AUTO
XX%)
function
(see
12. Press
[ESCAPE]
to
return
to
the
Auto
Cal
menu.
Figure 2
‐
15
The
gate 1
start
13. Press
Cal
‐
Vel
.
14. Use
the
knob
or
arrow
keys
to
adjust
the
displayed
value
to
the
appropriate
thickness.
For
this
example,
the
transducer
is
coupled
to
a
12.5 mm
step
(see
Summary of Contents for EPOCH 600
Page 1: ...DMTA 10007 01EN U8778365 Revision B June 2010 EPOCH 600 Basic Operation Manual ...
Page 6: ...DMTA 10007 01EN U8778365 Rev B June 2010 vi List of Abbreviations ...
Page 18: ...DMTA 10007 01EN U8778365 Rev B June 2010 12 Important Information Please Read Before Use ...
Page 32: ...DMTA 10007 01EN U8778365 Rev B June 2010 26 Chapter 1 ...
Page 56: ...DMTA 10007 01EN U8778365 Rev B June 2010 50 Appendix ...
Page 60: ...DMTA 10007 01EN U8778365 Rev B June 2010 54 List of Tables ...