33
100μsec. or less
Trigger input
Exposure
Readout &
transfer to PC
6
EXTERNAL TRIGGERING
When the cellSens / OLYMPUS Stream software is used, the DP73 can record still images or control a commer-
cially available shutter based on an external trigger signal.
Trigger Input
·
The trigger input from external equipment can be used to start still image acquisition using the cellSens /
OLYMPUS Stream.
·
The trigger input is detected at the edge and the cellSens / OLYMPUS Stream starts to acquire the images
based on the preset exposure time. The positive and negative logics can be switched with the cellSens /
OLYMPUS Stream.
·
The exposure is started within 100 μsec after the trigger input.
·
The trigger input is a TTL compatible signal.
V
IH
: 2.0 V (min.). V
IL
: 0.8 V (max.).
Trigger Input Timing Chart (Positive logic)