31
CX43
Objective
Magnification
Numerical aperture of
objective
NA
Numerical aperture of condenser suitable for
observation of specimen for microscope
(when setting to 70% of objective)
Plan objective
4X
10X
20X
40X
100XO
0.10
0.25
0.40
0.65
1.25
0.10 x 0.7 = 0.07
0.25 x 0.7 = 0.175
0.40 x 0.7 = 0.28
0.65 x 0.7 = 0.455
1.25 x 0.7 = 0.875
The aperture diaphragm is a diaphragm to adjust the numerical aperture of the condenser.
Adjusting the numerical aperture of the condenser appropriately with respect to the numerical aperture of each objective
allows you to observe the specimen with the best contrast and resolution.
Numerical aperture of objective and numerical aperture of condenser suitable for observation
In general, since the contrast of the specimen to be observed
with microscope is low, it is appropriate to adjust the
numerical aperture of the condenser to approximately 70 to
80% of the numerical aperture of the objective.
The rough indication is that the aperture diaphragm image
viewed when looking into the tube without eyepieces looks as
the picture shown on the right.
Aperture diaphragm image
Pupil of objective
(bright circular ring)
70-80%
30-20%
Aperture diaphragm