background image

Solution CIX Height Measurement

Installation Manual

4

1

CIX Height Measurement Solution

This manual describes the installation of the objective as well as the 
installation and configuration of the software for the CIX Height 
Measurement Solution. Read this manual carefully before installing and 
configuring the system. 

1.1

Safety instructions

1.2

Installing the objective

1. Use the joystick to move the stage all the way forwards to the 

mounting position. 

2. Remove the multi-sample holder.
3. Lower the stage to its lowest position by rotating the coarse focus 

adjustment knob (1).

Caution

Pinching hazard

When the stage moves, gaps appear. This creates a pinching hazard.
Make sure that you are not within the stage’s range of movement 
when it is moving.
Try never to put your hands or fingers into any gaps.

Summary of Contents for CIX100

Page 1: ...Installation Manual CIX Height Measurement Solution OLYMPUS Cleanliness Inspector System Optical Microscope and Accessories English ...

Page 2: ...the right to make such changes without obligation to notify the purchaser In no event shall Olympus Soft Imaging Solutions GmbH be liable for any indirect special incidental or consequential damages arising out of purchase or use of this manual or the information contained herein No part of this document may be reproduced or transmitted in any form or by any means electronic or mechanical for any ...

Page 3: ...urement Contents Installation Manual 1 CIX Height Measurement Solution 4 1 1 Safety instructions 4 1 2 Installing the objective 4 1 3 Installing the Solution 8 1 4 Configuring the objective 12 1 5 Calibrating the objective 13 ...

Page 4: ...figuring the system 1 1 Safety instructions 1 2 Installing the objective 1 Use the joystick to move the stage all the way forwards to the mounting position 2 Remove the multi sample holder 3 Lower the stage to its lowest position by rotating the coarse focus adjustment knob 1 Caution Pinching hazard When the stage moves gaps appear This creates a pinching hazard Make sure that you are not within t...

Page 5: ...Installation Manual 5 4 Use the joystick to move the stage all the way back towards the frame 5 Loosen the screw that holds the nosepiece using the hexagonal screwdriver To do that use the screw that is located at the microscope frame ...

Page 6: ...jective from its packaging 10 Screw the objective in to position 4 11 CIX90 The following steps should be performed by two people if possible Connect the port of the nosepiece with the plug of the cable under the microscope frame Place the nosepiece back in the runner on the microscope While you do this keep the cable out of the way by pulling it carefully back towards the back of the microscope f...

Page 7: ... Installation Manual 7 12 CIX100 Place the nosepiece in the runner on the microscope and push it slowly all the way to the back 13 When the nosepiece is all the way back on the runner tighten the screw with the hexagonal screwdriver ...

Page 8: ...to your DL DVD drive The start dialog box of the installation program should be automatically displayed on your screen Note It is possible that the autorun function of your operating system is deactivated In this case double click the setup exe file located in the root directory of your DL DVD drive The installation program s first dialog box asks you to enter the license key for the solution Plea...

Page 9: ...net connection The activation takes place in the background Click the Contact Server button to complete the activation of the OLYMPUS Cleanliness Inspector Software Note If you are not connected to the Internet but are able to exchange files with another computer that does have an Internet connection please follow the steps in the next dialog box ...

Page 10: ...Solution CIX Height Measurement Installation Manual 10 The following website is used for the activation ...

Page 11: ... another computer that does have an Internet connection you can activate the OLYMPUS Cleanliness Inspector Software by telephone Call our Customer Service Center and follow the instructions below The activation is now complete You can use your software on the computer for which it was activated even if it doesn t have an Internet connection ...

Page 12: ...al Motorized BX3M CB entry in the tree view Configuration in the Device Settings dialog box 5 Select position 4 in the Pos column 6 Select the 20 entry in the Magnification list 7 Select the MPLFLN entry in the Objective Type list 8 Enter the value 500 μm in the Escape Distance field 9 In the Escape group select the Auto Escape check box This prevents the objective from colliding with the sample w...

Page 13: ...m the following calibration processes for the newly installed objective Stage Limits XY Objective Shift Parfocality Shading Correction and White balance Manual Magnification Calibration Manual Z Axis Calibration It s not necessary to recalibrate the objectives that have already been calibrated For more information on the calibration processes see the software s help document ...

Page 14: ... 48 Woerd Avenue Waltham MA 02453 U S A OLYMPUS MEDICAL SYSTEMS INDIA PRIVATE LIMITED Ground Floor Tower C SAS Tower The Medicity Complex Sector 38 Gurgaon 122001 Haryana INDIA OLYMPUS KOREA CO LTD 8F Olympus Tower A 446 Bongeunsa ro Gangnam gu Seoul 135 509 Korea OLYMPUS AUSTRALIA PTY LTD 3 Acacia Place Notting Hill VIC 3168 Australia OLYMPUS SINGAPORE PTE LTD 491B River Valley Road 12 01 04 Vall...

Reviews: