Problem
Cause
Remedy
Page
e) Coarse adjustment will not go all
the way down.
Condenser holder is too low.
Raise condenser holder.
12
f ) Objective makes contact with speci-
men before focus is obtained.
Specimen is mounted upside down.
Mount specimen correctly.
–
4. Observation Tube
a) Field of view of one eye does not
match that of the other.
Interpupillary distance is incorrect.
Adjust interpupillary distance.
11
Incorrect diopter adjustment.
Adjust diopter.
11,22
Different eyepieces are used on left
and right.
Change on eyepiece to match the
other so that both sides are the same
type.
–
Your view is not accustomed to micro-
scope observation.
Upon looking into eyepieces, try
looking at overall field before concen-
trating on specimen range. You may
also find it helpful to look up and into
distance for a moment before looking
back into microscope.
–
5. Stage
a) Image shifts when you touch
stage.
Stage is not properly mounted.
Clamp stage.
21
b) Specimen stops midway on the
X-axis traverse.
Specimen is not correctly positioned.
Place specimen correctly.
19
c) Horizontal and/or X- and Y-axis kn-
obs are too tight or too loose.
Tension of horizontal and/or X- and
Y-axis knobs is too high or too low.
Adjust tension.
20
d) Stroke has reduced.
Stage guide is deviated.
Use treatment method described on
page 20.
20