18
BX51/52-P
Fig. 23
(5) Clamp the stage with the stage rotation clamping knob.
When Using Polarizing Intermediate Attachment (U-CPA)
· Normal Observation
# Make sure the 10X objective is inserted into the primary hole on the
centerable revolving nosepiece.
1. Pull out the Bertrand lens knob @ to the OUT (¦) position. (Fig. 23)
2. Center the condenser. (Refer to page 18 in the BX51/52 instruction manual.)
3. Follow steps 9 and 10 in “When Using the Intermediate Attachment
for AN360".
‡
Fig. 22
10. This concludes centering of the optical axis for the 10X objective, which
will act as the reference objective. Now, center the other objectives with
the centerable nosepiece by engaging the objectives into the light path
one after another.
@
Summary of Contents for BX51
Page 33: ...MEMO ...