18
BX53/51/41-P
Fig. 23
(5) Clamp the stage with the stage rotation clamping knob.
When Using Intermediate Attachment for conoscopic and orthoscopic
observation (U-CPA)
· Normal Observation
Make sure the 10X objective is inserted into the primary
hole on the centerable revolving nosepiece.
1. Pull out the Bertrand lens knob @ to the OUT (¦) position. (Fig. 23)
2. Center the condenser. (For details, see instructions of each microscope.)
3. Follow steps 9 and 10 in “When Using the Intermediate Attachment
for orthoscopic observation".
‡
10. This concludes centering of the optical axis for the 10X objective, which
will act as the reference objective. Now, center the other objectives with
the centerable nosepiece by engaging the objectives into the light path
one after another.
@
CAUTION
Fig. 22