 : Turns ON.   : Blinks.   :Turns OFF.   : Varies depending on state.

Phenomena

Cause

Remedy

Page

Number 

of beeps

CB

HS

MIX

5 times

The 

communication 

between the control 

box (BX3M-CBFM 

or BX3M-CB) and 

the hand switch 

for motorized 

nosepiece 

(BX3M-HSRE) is 

disconnected.

The cable of 

the hand switch 

for motorized 

nosepiece (BX3M-

HSRE) is removed 

while the power 

is ON.

Set the main switch to   (OFF) 
once and set the main switch to   
(ON) again.
If the error occurs frequently, contact 
Olympus.

9

5 times

The 

communication 

between the 

control box (BX3M-

CBFM or BX3M-

CB) and PC is 

disconnected.

Set the main switch to   (OFF) 
once and set the main switch to   
(ON) again.
If the error occurs frequently, contact 
Olympus.

9

Summary of Contents for BX3M-CB

Page 1: ...of this product we recommend that you study this manual thoroughly before operating this product and always keep this manual at hand when operating this product Retain this instruction manual in an ea...

Page 2: ...ted in a commercial environment This product generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interferenc...

Page 3: ...reflected light observation 10 2 Turning ON the illumination of the MIX slider for reflected light observation 11 3 Adjusting the brightness 12 4 Selecting the illumination pattern 13 2 3 Operating t...

Page 4: ......

Page 5: ...M CB and BX3M CBFM For usage in combination with microscope refer to the instruction manual for your microscope The control box BX3M CB and BX3M CBFM are used to control motorized revolving nosepiece...

Page 6: ...ury NOTE Indicates a potentially hazardous situation which if not avoided may result in damage to the product or other property TIP Indicates the useful knowledge or information for use CAUTION Instal...

Page 7: ...he power cord and that of the power outlet If the product is not grounded our intended electric safety and EMC performance of the product can not be assured Do not use the product in close proximity t...

Page 8: ...an excessive force the control box may drop and you may get injured The following symbols are placed on the product Study the meaning of the symbols and always use the product in the safest possible...

Page 9: ...ctive for a long time since it may cause damage to your eyes Do not expose your skin to the light coming out from the objective for a long time If your skin is exposed to the light coming out from the...

Page 10: ...is a precision instrument Handle it with care and avoid subjecting it to a sudden or severe impact Never disassemble any part of the product Otherwise failure could be caused Do not use this product i...

Page 11: ...AC adapter jack Connector for USB cable Connector for motorized nosepiece Connector for hand switch for motorized nosepiece exposure Connector for hand switch for motorized nosepiece exposure Connecto...

Page 12: ...ith the hand switch BX3M HSRE Operating portions for MIX slider for reflected light observation Function buttons Arbitrary functions can be assigned when connecting to the PC Operating portions for mo...

Page 13: ...e main switch of BX3M CBFM to ON The beep sound is heard once 1 1 This section describes the procedure to turn ON the main switch of the BX3M CBFM If you use the BX3M CB turn ON the main switch of the...

Page 14: ...nd is heard MIX slider for reflected light observation position Light path First level pulled out OUT Second level pressed in IN 1 1 Inserting the MIX slider for reflected light observation The state...

Page 15: ...set it to ON the illumination of the MIX slider for reflected light observation U MIXR turns ON State Indicator b c Function ON Turns ON The illumination turns ON OFF Turns OFF The illumination turns...

Page 16: ...predetermined quantity Long press Darken continuously Short press Brighten by predetermined quantity Long press Brighten continuously Indicator The light intensity state can be checked by the indicato...

Page 17: ...pattern Operation Function Short press Changes the illumination pattern Long press Short press while the illumination pattern is rotating automatically The illumination pattern rotates automatically c...

Page 18: ...n position The illumination position shifts in the rotated direction by one scale at a time Example of clockwise rotation While the illumination pattern is rotating automatically you cannot operate th...

Page 19: ...pressing the button can be changed by setting the dip switch of the control box BX3M CBFM or BX3M CB For details see 5 3 Setting the dip switch page 33 Before setting the dip switch set the main switc...

Page 20: ...cable length 2 m RS 232C cable length 2 m External interface RS 232C external cable is included D SUB 9 pin female straight connector USB 2 0 external cable is included Operating environment Indoor us...

Page 21: ...ate Phenomena Cause Remedy Page Number of beeps CB HS MIX a The motorized nosepiece does not rotate even though pressing the button of the hand switch for motorized nosepiece BX3M HSRE None The contro...

Page 22: ...e hand switch for motorized nosepiece BX3M HSRE is not connected Set the main switch to OFF and connect the hand switch for motorized nosepiece BX3M HSRE If the indicator4 on the side or the back of t...

Page 23: ...t Set the main switch to OFF and use the dip switch on the side of the control box BX3M CBFM or BX3M CB to set the number of objective mounting holes on the nosepiece 9 33 The motorized nosepiece did...

Page 24: ...ted Set the main switch to OFF and connect the control box and the AC adapter of the control box again Then set the main switch to ON and confirm that the beep is heard once 9 27 28 The control box BX...

Page 25: ...back of the control box does not light even though the hand switch BX3M HS is connected contact Olympus 9 27 None Turns OFF The MIX slider for reflected light observation U MIXR is not connected Conn...

Page 26: ...he MIX slider for reflected light observation in the light path If the POSITION indicator of the hand switch BX3M HS keeps lighting OUT IN does not light even though the MIX slider for reflected light...

Page 27: ...act Olympus 12 d The OB indicator does not light even though selecting the objective by turning the motorized nosepiece None The control box BX3M CBFM or BX3M CB is not connected Set the main switch t...

Page 28: ...iver is not installed Install the device driver from the installation media of the application software The device is not detected properly Scan the device again with the Windows device manager The US...

Page 29: ...caused by thunder etc and the system was reset This is not a failure If this phenomenon occurs frequently contact Olympus g The beep sound is heard five times 5 times The communication between the co...

Page 30: ...disconnected The cable of the hand switch for motorized nosepiece BX3M HSRE is removed while the power is ON Set the main switch to OFF once and set the main switch to ON again If the error occurs fre...

Page 31: ...nnecting cables set the main switch to OFF and unplug the power cord from the outlet Cables are vulnerable when bent or twisted Never subject them to excessive force Connect the connectors in the corr...

Page 32: ...adapter jack b of the FM control box When inserting or removing the AC adapter plug hold the plug and insert straightly 1 NOTE Insert the power cord connector c in the connector d of the AC adapter 2...

Page 33: ...29 BX3M CB CBFM e f 3 Connect the power cord plug e to the power outlet f For the safety purpose connect the power cord plug last 3 NOTE...

Page 34: ...terface cables provided by Olympus If you use the commercially available USB 2 0 cables or hubs the system operations cannot be ensured Be sure to connect cables when the main switch of BX3M CBFM or t...

Page 35: ...the other connector of the USB cable to the USB 2 0 connector of PC Some PCs are equipped with both USB 2 0 and USB 3 0 connectors Be sure to connect the USB cable to the USB 2 0 connector If it is co...

Page 36: ...32C connector of PC 1 2 a c 1 b 1 BX3M CB BX3M CBFM a 2 Start procedures Exit procedures Start procedures Turn ON the power of PC and log on Windows Wait until the desktop appears Set the main switch...

Page 37: ...side of BX3M CB 1 1 1 BX3M CBFM BX3M CB Factory default setting Function Unit Setting 1 2 3 4 Beep OFF The beep sound is heard ON The beep sound is not heard Motorized nosepiece OFF 5 holes ON 6 holes...

Page 38: ...on IEC configuration appliance coupling Voltage rating Current rating Temperature rating Length Fittings configuration Specifications A power supply cord should be certified by one of the agencies lis...

Page 39: ...tification mark Argentina IRAM Italy IMQ Australia SAA Japan JET JQA Austria VE Netherlands KEMA Belgium CEBEC Norway NEMKO Canada CSA Spain AEE Denmark DEMKO Sweden SEMKO Finland FEI Switzerland SEV...

Page 40: ...0 30 10 Verband Deutscher Elektrotechniker VDE e V Pr fstelle VDE HAR 30 10 10 Union Technique de I Electricite UTE USE HAR 30 10 30 Instituto Italiano del Marchio di Qualita IMQ IEMMEQU HAR 10 30 50...

Page 41: ...AR 30 30 50 Norges Elektriske Materiellkontroll NEMKO NEMKO HAR 10 10 70 Asociacion Electrotecnica Y Electronica Espanola AEE UNED HAR 30 10 70 Hellenic Organization for Standardization ELOT ELOT HAR...

Page 42: ...38 MEMO...

Page 43: ......

Page 44: ...Distributed by Manufactured by Shinjuku Monolith 2 3 1 Nishi Shinjuku Shinjuku ku Tokyo 163 0914 Japan www olympus global com AX9451 01...

Reviews: