SPECIFICATIONS
Item
Specification
Vertical illuminator
Universal Reflected Illuminator
BX3-URA
Coded Fluorescence Illuminator
BX3-RFAS
· UIS2 optical system
· Super-widefield: NA 26.5
· Widefield: NA 22
· Observation switching : Mirror unit turret carrying max. 8 mirror units.
· Aperture iris diaphragm and field iris diaphragm
(Both centerable)
· Aperture iris diaphragm and field iris diaphragm
(Aperture iris diaphragm centerable)
· Shutter provided
· Slider insertion slots
@
Analyzer
2
Polarizer
3
ND filters
@
Analyzer
2
ND filters (using the provided ND filter slider)
· Available observation modes
@
Reflected fluorescence
2
Reflected fluore Transmitted DIC
3
Reflected fluore Phase contrast
| Transmitted light
· Optional accessories
U-CBS information readout control box:
For acquisition of mirror unit position information
Lamp housing for mercury burner · Lamp housing for 100 W mercury burner U-LH100HG
· Apo lamp housing for 100 W mercury U-LH100HGAPO
· Mercury burner: USH-103OL (OLYMPUS)
· Power supply unit U-RFL-T
Mercury burner
· DC, 100 W, ultrahigh-pressure mercury burner
· USH-103OL
*
(OLYMPUS)
· Life: 300 hours (2-hour ON/30-min. OFF cycle)
Operating environment
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104° F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly through
70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104° F).
· Supply voltage fluctuations; Not to exceed ±10% of the normal voltage.
· Pollution degree: 2 (in accordance with IEC60664-1)
· Installation/Overvoltage category: II (in accordance with IEC60664-1)
*
The mercury burner, such as USH-102D (mfd. by Ushio) or HBO103W/2 (mfd. by OSRAM) can
also be used, but their performances are not guaranteed.