DMTA-10045-01EN, Rev. E, August 2016
Chapter 6
158
12. Adjust the
GAIN
(B key) to set the highest dot at approximately 90 % of the
screen height (see Figure 6-51 on page 158).
Figure 6
‑
51 The GAIN adjusted to set the highest dot
13. Press the ALARM menu key (
) twice to display the
XY ALM 1
screen, and
then set the
BOTTOM
(C key) to
30 %
.
14. Press the FULL NEXT key (
) to display the full screen, and then slowly scan
over the disbonds and ensure that the dots still match the signal (see Figure 6-52
on page 158).
Figure 6
‑
52 The second scan over the disbonds