DMTA-10045-01EN, Rev. E, August 2016
Applications
139
Figure 6
‑
23 The Smaller Disbonds and Repair Identification application
To calibrate the signals
1.
Press the MAIN menu key (
) once, and then set the
FREQ
(A key) to
10 kHz
using the knob.
2.
Place the probe tip on a good part of the standard (CFRP side), and then press the
CAL NULL key (
).
3.
Slowly scan over one of the 13 mm (0.5 in.) disbonds, and then press the FREEZE
key (
) [see Figure 6-24 on page 139].
Figure 6
‑
24 The scan signal of a disbond