DMTA-10045-01EN, Rev. E, August 2016
Applications
125
3.
Set the probe tips over a good area of the standard, and then press the CAL NULL
key (
).
4.
Scan over the far-side and near-side disbonds, and then, while you keep the probe
moving over the disbonds, ensure that both defects are detected (see Figure 6-4 on
page 125).
Figure 6
‑
4 Detecting the far
‑
side and near
‑
side disbonds
GATE setup
By default, the
GATE
is set to
AUTO
. In
AUTO
mode the BondMaster 600
automatically detects the peak signal from the RF view and uses it to construct the XY
flying dot view.
5.
If desired, manually set the gate to the desired position by pressing
GATE
(D key)
and rotating the knob.
A recommended
GATE
position would be the peak of the first reflection.
The recommended
GATE
position is often to the left of the first strongest signal
peak (see Figure 6-5 on page 126).
NOTE
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...