BondMaster 600 — Getting Started Guide

5

To Start Up

When power is turned on, the BondMaster 600 starts up 
in one of two modes, depending on what is connected to 
the instrument.

If no probe is connected, or if a probe that is not a 
PowerLink type of probe is connected, the first screen 
displayed by the BondMaster 600 software is the 
application quick-setup menu. From this menu, choose 
a basic application to automatically configure the 
appropriate instrument settings.

The BondMaster 600 applications are designed for quick setup of the instrument. However, always 
follow published maintenance procedures when inspecting.

To Navigate the Application Menu

Navigating the application menu is designed to be intuitive, and the settings for each application are 
designed to permit immediate inspection; little or no additional instrument setup is needed.

1.

Rotate the knob to highlight one of the applications.

2.

Press the Enter key (

) to select the application.

OR

Press the Return key (

) to go back to the BondMaster 600 main screen.

Alternatively, if a PowerLink probe is connected to the instrument before it is turned on, the 
BondMaster 600 starts up in the PowerLink recognition screen, and you can proceed as follows:

In the PowerLink recognition screen, load the 
program stored in the PowerLink probe by pressing 
the A key. This automatically sets up the 
instrument.
OR
Bypass the stored program in the probe and gain 
access to the instrument’s main inspection screen by 

pressing the Return key (

).

NOTE

Summary of Contents for BondMaster 600

Page 1: ...n imminently hazardous situation and calls attention to a procedure practice or the like which if not correctly performed or adhered to will result in death or serious personal injury Indicates a pote...

Page 2: ...51431 8 cell battery holder with connector plug Olympus P N 600 BAT AA U8780295 Pitch and catch and MIA mode cable for BondMaster 600 3 3 meter 11 foot length 11 pin to 11 pin connectors Olympus P N S...

Page 3: ...ding alkaline batteries in the battery holder P N 600 BAT AA U8780295 or use any other charger adaptor Doing so might cause an explosion or injury Do not attempt to power or charge other electronic eq...

Page 4: ...thumb screws to complete the installation To Install the MicroSD Card 1 Remove the MicroSD card P N MICROSD ADP 2GB U8779307 from its packaging 2 Loosen the two thumb screws and then open the BondMas...

Page 5: ...on Menu Navigating the application menu is designed to be intuitive and the settings for each application are designed to permit immediate inspection little or no additional instrument setup is needed...

Page 6: ...aximum of two real time readings from among the available choices You can set the real time readings bar to display one or two readings or you can deactivate it The instrument settings are displayed o...

Page 7: ...eturn to a previous screen CAL NULL If this direct function key is pressed once it nulls zeros the BondMaster 600 If this key is pressed and held a calibration wizard is initiated in RESON resonance a...

Page 8: ...isplay MAIN Provides access to the main menu which controls functions such as frequency gain angle filters RF signal and gate DISP DOTS Provides access to the display menu which controls functions suc...

Page 9: ...cell contents which are corrosive and explosive Do not short circuit the battery terminals A short circuit could cause injury and severe damage to a battery making it unusable Do not expose a battery...

Page 10: ...Fischer connector Up to 80 V can be present on the inner conductors The warning symbol shown in the figure below warns of this electric shock risk The instrument must only be connected to a power sour...

Page 11: ...BondMaster 600 Getting Started Guide 11...

Page 12: ...Copyright 2014 by Olympus All rights reserved All brands are trademarks or registered trademarks of their respective owners and third party entities Part ID U8030413 U8030413 Printed on Rolland Hitec...

Reviews: