10-015490-01EN, Rev. 2, June 2020
Chapter 2
32
Figure 2-2 on page 32 shows the LEDs described in Table 5 on page 31. Table 6 on
page 33 describes the behavior of the LEDs in relation to X-ray emissions.
Figure 2-2 BTX III LED indicators
2
High voltage/X-ray On LEDs
These LEDs glow red whenever high voltage is enabled and X-ray
generation is possible.
3
Interlock open LED
This LED is normally off. It glows amber when a safety interlock
switch has been opened. A safety interlock switch could be triggered
by removing the sample carrier during a sample analysis.
4
Power key switch
The power key switch has
ON
and
OFF
positions. The key must be
inserted and in the
ON
position for the system to operate, including
operation of the X-ray beam.
5
Control panel interlock
(not shown)
This is a magnetic switch that prevents the operation of high voltage
when the control panel (top panel) is improperly seated or removed
from the outer case. Under these conditions, the X-ray beam will not
operate.
Table 5 BTX III radiation safety features
(continued)
Label
Safety feature
High Voltage/X-ray On LEDs
Interlock open LED
1
2
Summary of Contents for BIX III
Page 6: ...Table of Contents vi 10 015490 01EN Rev 2 June 2020...
Page 8: ...10 015490 01EN Rev 2 June 2020 List of Abbreviations viii...
Page 12: ...10 015490 01EN Rev 2 June 2020 Labels and Symbols 4...
Page 28: ...10 015490 01EN Rev 2 June 2020 Introduction 20...
Page 36: ...10 015490 01EN Rev 2 June 2020 Chapter 1 28...
Page 64: ...10 015490 01EN Rev 2 June 2020 Chapter 4 56...
Page 82: ...10 015490 01EN Rev 2 June 2020 Chapter 5 74...
Page 84: ...10 015490 01EN Rev 2 June 2020 Appendix A 76...
Page 100: ...10 015490 01EN Rev 2 June 2020 List of Tables 92...