background image

Complete Teardown, Cleaning, and Reassembly of the Olympus BH2-6RE Modular Revolving Nosepiece               

   

 

Page 15 of 18 

 

 

Revision 3 

outer  perimeter  of  the  revolving  turret,  spaced 
approximately 180° apart (see 

Figure 42

). 

 

Figure 42 

 Apply black adhesive to the revolving turret 

Carefully reinstall the knurled-rubber grip ring onto the 
outer perimeter of the revolving turret, making sure to 
not stretch or damage the grip ring (see 

Figure 43

). 

 

Figure 43 

 Reinstall the knurled-rubber grip ring 

Use dry cotton swabs to thoroughly remove any visible 
silicone RTV squeeze-out (see 

Figure 44

). 

 

Figure 44 

 Remove any silicone RTV squeeze-out 

Ready for Service 

The  newly  reconditioned  BH2-6RE  modular  revolving 
nosepiece assembly is now ready to provide many more 
years of service (se

Figure 45

)

 

Figure 45 

 BH2-6RE ready to be put back into service 

Requirements for Periodic Maintenance 

Periodic cleaning and application of fresh grease to the 
mechanical  detent  notches  in  the  stationary  base  is 
necessary to minimize wear of the mechanical detents, 
thereby  maximizing  the  useful  service  life  of  the 
BH2-6RE.    This  can  be  easily  accomplished  by  simply 
removing the protective cover, cleaning and re-greasing 
the detent notches, and then reinstalling the protective 
cover  per  the  procedures  detailed  in  this  document.  
The  dovetail  slide  does  not  need  to  be  removed  to 
perform  this  periodic  maintenance.    If  the  equipment 
sees heavy usage, this service should be performed on a 
six-month interval

3

.   

Problems with the Turret Assembly 

A  few  problems  with  the  turret  assembly  can 
sometimes be found in the reassembled nosepiece.  The 
first  will  be  seen  if  one  or  more  of  the  mechanical 
detents are excessively worn.  This will cause radial float 
of  the  revolving  turret  in  one  or  more of  the  objective 
positions,  and  these  objectives  will  have  trouble 
returning  to  and  maintaining  their  proper  index 
position.    This  can  make  it  difficult  to  utilize  some 
illumination  types,  such  as  phase  contrast,  since  the 
phase  annuli  will  not  be  able  to  hold  an  acceptable 
alignment due to variations in objective indexing. 
The  other  problem  that  may  be  seen  is  caused  by 
overall  float  of  the  revolving  turret,  relative  to  the 
center  point  of  the  stationary  base.    If  there  is  excess 

 

3

  Field  experience  has  shown  that  even  heavily  used  scopes  (such  as  those 

used  in  hospitals  and  clinical  lab  settings),  when  lubricated  on  a  six-month 
interval, can be expected to provide many years of trouble-free service. 

Summary of Contents for BH2-6RE

Page 1: ...tribute this PDF document in its entirety for personal non commercial purposes only provided that the contents are not modified in any way including the copyright notices contained herein This documen...

Page 2: ...rimeter Bearing Balls 10 Remove the Stationary Base from the Turret 10 Clean Grease from the Various Components 11 Reinstall Stationary Base into Revolving Turret 11 Apply Ring of Grease for the Perim...

Page 3: ...ure 18 Components of the turret assembly 10 Figure 19 Loosen the threaded retaining ring 10 Figure 20 Remove the threaded retaining ring 10 Figure 21 Remove the perimeter bearing balls 10 Figure 22 Li...

Page 4: ...f 18 Revision 3 Figure 43 Reinstall the knurled rubber grip ring 15 Figure 44 Remove any silicone RTV squeeze out 15 Figure 45 BH2 6RE ready to be put back into service 15 Figure 46 Revolving turret w...

Page 5: ...on revolving modular nosepiece assembly used on the Olympus BHS BHSP BHT and BHTP microscope stands This procedure is also applicable the BH2 5RE five position modular revolving nosepiece assembly due...

Page 6: ...due to solvent absorption The list of solvents generally considered safe for nylon includes acetone diethyl ether heptane mineral spirits naphthalene and turpentine Before Starting the Overhaul Remov...

Page 7: ...stment screw see Figure 8 can be difficult to remove unless the proper tool is used Do not attempt to remove the slotted lock ring and pivot adjustment screw unless you have access to such a tool sinc...

Page 8: ...ds in the revolving turret Figure 10 Insert screwdriver into bores to lock turret Loosen and Remove the Slotted Lock Ring Hold the turret assembly by grasping the knurled rubber grip ring on the outer...

Page 9: ...lace with adhesive and can be surprisingly stubborn to remove so be sure to use the proper JIS screwdriver to prevent damaging the screw heads It might also be helpful to heat the screws with a heat g...

Page 10: ...ter punch or nailset tool into one of the six detent notches in the revolving turret instead of into one of the four tooling holes in the threaded retaining ring or the revolving turret will be irrepa...

Page 11: ...er of the revolving turret see Figure 24 Do not apply too much grease here to minimize the squeeze out that will occur when the threaded retaining ring is reinstalled to hold the stationary base into...

Page 12: ...that the center pivot bore has been greased reinstall the bearing ball into the freshly greased center pivot bore The grease will hold the center pivot bearing ball in the proper position during subs...

Page 13: ...oosen the slotted lock ring readjust the pivot adjustment screw and retighten the slotted lock ring as described above until the turret motion feels acceptable Clean Off Any Visible Grease Use a suita...

Page 14: ...the stationary base see Figure 40 Be sure to orient the dovetail slide such that the relief notch see inset of Figure 40 is facing downwards and towards the slotted lock ring on the pivot adjustment s...

Page 15: ...removing the protective cover cleaning and re greasing the detent notches and then reinstalling the protective cover per the procedures detailed in this document The dovetail slide does not need to be...

Page 16: ...n properly adjusted to remove the play in the center pivot mechanism How to Contact the Author Please feel free to direct any questions or comments regarding this document or BH 2 microscopes in gener...

Page 17: ...resistant layer of dull silver colored plating on the exposed surface When the hard plating in the area of the mechanical detents begins to wear away as a result of severe use without routine cleaning...

Page 18: ...t Item Description Manufacturer Manufacturer Model Vendor Vendor Price S1 E6000 adhesive black 2 oz 32329 Amazon 8 22 S2 Plastilube Brake Grease 75 cc Plastilube ATE70015 Amazon 10 88 Plastilube Brake...

Reviews: