DMTA-10022-01EN, Rev. C, January 2015
Chapter 6
74
a)
In the measurement screen, press
[SETUP]
, and then select
MEAS
.
b)
In the
MEAS
menu, set the
MEAS OPTION
to
MANUAL E-TO-E
, and then
press
[MEAS]
.
2.
Press
[GAIN/WAVE ADJ]
.
The wave adjustment parameter appears (see Figure 6-5 on page 74).
Figure 6-5 Comparing manual measurements
3.
Use the
[
]
and
[
]
keys to select either the
EXT BLANK
or the
E1 BLANK
parameter.
4.
Use the
[
]
and
[
]
keys to adjust the value to exclude unwanted peaks and to
detect the desired echoes.
6.2.2
Dual Element Transducer Selection in Echo-to-Echo Modes
Although the optional Echo-to-Echo modes work with all the 45MG dual element
transducers, Olympus recommends using specific transducers for particular thickness
ranges in steel parts (see Table 4 on page 75).
Poor manual measurement
Good manual measurement
[WAVE ADJ]
parameter
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...