DMTA-10022-01EN, Rev. C, January 2015
Custom Setups for Single Element Transducers
171
Figure 10-5 Mode 3 detection example
Refer to “Echo Window” on page 177 for information on the modes in relation with
the echo window.
10.4 First Peak
With single element transducers, the 45MG normally detects peaks on either the
highest positive or highest negative peak of the RF waveform. This feature works well
for most precision thickness applications.
This normal peak detection can be unstable for applications in which the back-wall
echo is irregular and contains several negative or positive peaks that are close in
amplitude. In such cases, the thickness reading can fluctuate as the gage switches
detection from one peak to another. This can happen, for example, in applications
measuring the length of bolts, or the thickness of the gel coat over fiberglass (see
Figure 10-6 on page 172). In such cases, help stabilize the echo detection and thickness
measurements by selecting the first peak algorithm to detect the first of several peaks
of similar amplitudes (see step 5.
in “Creating a Custom Setup for a Single Element
Interface echo
Second back-wall
echo
First back-wall echo
Detection bar
Mode indicator
NOTE
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...