DMTA-10022-01EN, Rev. C, January 2015
Custom Setups for Single Element Transducers
165
10. Custom Setups for Single Element Transducers
Once the Single Element High Resolution option has been activated, the 45MG
includes predefined setups for standard single element transducers. In some cases, a
45MG ships from the factory preprogrammed with one or more custom setups to
meet special customer requirements. You can create your own custom setups to meet
the need of a particular single element transducer, or for a particular application.
Predefined and custom setups allow you to quickly change between transducer and
application setups by selecting a previously saved setup.
10.1 Creating a Custom Setup for a Single Element Transducer
A custom setup should be created when the measurement requirements of a
particular application are not optimally met by one of the standard setups. Once the
adjustments are made, you can name and store this setup in one of the 35 user-defined
setup locations.
The adjustments described in the following procedure and subsections should be
made only by a qualified technician who is familiar with the basic theory of ultrasonic
gaging and the interpretation of ultrasonic waveforms.
Many of the adjustments that can be made in a setup are interactive. All adjustments
have an effect on the measurement range and/or measurement accuracy of the 45MG.
In most cases, do not attempt adjustments without monitoring the waveform.
CAUTION
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...