Configuring System Parameters
The
SYSTEM
screen, allows you to configure many 45MG system parameters.
To configure system parameters
1.
In the measurement screen, press
[SETUP]
, and then select
SYSTEM
.
The
SYSTEM
screen appears (see Figure 8-2 on page 144).
Figure 8-2 The SYSTEM screen
2.
Set
BEEPER
to
ON
or
OFF
(see “Setting the User Interface Language and Other
System Options” on page 43 for details).
3.
Set
INACTIVE TIME
to
ON
or
OFF
(see “Setting the User Interface Language
and Other System Options” on page 43 for details).
4.
In the
LANGUAGE
text box, select the desired user interface language (see
“Setting the User Interface Language and Other System Options” on page 43 for
details).
5.
Set
RADIX TYPE
to the desired character (period or comma) for separation of the
integer and decimal numbers.
6.
Set
SAVE KEY DATA
to save only the thickness measurement (
THICKNESS
), or
both the thickness and the waveform (
THI WF
).
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...