DMTA-10004-01EN, Rev. D, November 2016
Using Advanced Gaging Features
141
Figure 10
‑
5 The half
‑
size and full
‑
size B
‑
scans
B
‑
SCAN DIRECTION
:
Choose the B-scan direction to match the direction of transducer movement. A
scan direction arrow appears below the left corner of the B-scan display to
indicate the transducer scan direction (see Figure 10-6 on page 141). The data
starts appearing on the screen in the opposite direction.
LEFT TO RIGHT
The transducer scans the part from left to right and the data starts appearing
at the right of the screen and scrolls to the left.
RIGHT TO LEFT
The transducer scans the part from right to left and the data starts appearing
at the left of the screen and scrolls to the right.
Figure 10
‑
6 B
‑
scan elements
Half size B-scan
Full size B-scan
Scan direction
(left to right)
Min/Max marker
B-scan range
Loss-of-signal (LOS)
indicator
Data scrolls in from
right to left
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...