DMTA-10009-01EN [U8778346], Rev. D, November 2016
Chapter 5
72
You can also achieve a material sound velocity and zero calibration procedure by
performing the operations in the reverse order, starting with the zero calibration,
followed by the material sound velocity calibration.
5.2.5
Entering a Known Material Sound Velocity
When preparing to measure thicknesses on parts made of a different material, if you
know the sound velocity for the material, you can directly enter the velocity without
performing a material sound velocity calibration procedure.
To enter a known material sound velocity
1.
From the measurement screen, press
[2nd F]
,
[CAL VEL] (VEL)
.
2.
In the
VELOCITY
screen (see Figure 5-7 on page 72), use the arrow keys to edit
the velocity to the known value.
Figure 5
‑
7 Entering a known material sound velocity
3.
Press
[MEAS]
to return to the measurement screen.
NOTE