28
UNO-CONTROLLER can work in two alternative modes:
Chamber Control
mode (see paragraph 8.2.1)
and
Sample Control
mode (see paragraph 8.2.1.2).
Note
►
The Chamber Control mode is set by default.
8.2.1.1
Chamber Control mode
To select the Chamber Control mode, press on the Chamber icon (see pointer 1 in Figure 33 b) and press
on
Save
to confirm (see pointer 2 in Figure 33 b).
a
b
1
2
Figure 33. Settings
–
Control Mode
–
Chamber Control mode (a
–
b).
In Chamber Control mode, the temperature of the Chamber Base and Chamber Lid of Okolab H301
incubating chamber are strictly controlled to guarantee that the sample temperature is maintained at the desired
set point value.
Note
►
The Chamber Control mode affords a quick experiment start. No further actions are required for
multi user applications.
Tip
►
In Chamber Control mode, there is no need to place the T Sensor (the fine gauge thermocouple) into a
reference well during experiment.
After selecting Chamber Control mode, the screen in Figure 34 b will appear:
a
b
1
2
Figure 34. Settings
–
Control Mode
–
Chamber Control mode
–
Offset (a
–
b).
At this point you can choose:
1.
By pressing
Factory
button (see pointer 1 in Figure 34 b) UNO-CONTROLLER will recover and use the
factory calibration parameters.
Note
►
Factory calibration was performed at a Room Temperature of 23°C ±1°C for a temperature set point
of 37°C. Use factory calibration parameters if your Room Temperature is 23°C ±1°C.