39
6.
Press the Pump/Diffusion Mode button
(see Figure 49 a) to select the Aspiration Mode (see
Once you have followed the step above, LEO is ready to be used in Aspiration Mode. See paragraph 10.1.3
to start a Single Point Measurement or 10.1.4 to start a gas measurements logging.
Figure 49. How to select the Aspiration Mode (a
–
b).
10.1.3
CO2-O2 Single Point Measurement
To start a single point measurement, follow the instructions below:
1.
Select the Device for which the data will be stored (see paragraph 9.3.1).
Note
►
If no device is selected, the data will be stored on the current device shown on the Homepage (1 in
2.
Press the CO2-O2 Single Point Measurement icon
(see Figure 50 a). A pop-up dialog box appears
(see Figure 50 b). Select
Auto
for an automatic or
Manual
for a manual measurement.
−
Auto:
the CO2-O2 Single Point measurement terminates automatically when the gas reading
value becomes stable.
−
Manual:
you have to
click again on the CO2-O2 Single Point Measurement Icon to stop the
measurement.
Okolab recommends using the Auto measurement only when the gas flowrate sent to LEO is
higher than 130 ml/min.
Tip
►
Okolab suggests to use the Manual measurement to meet fully accuracy requirements.
a
b
Summary of Contents for LEO
Page 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Page 2: ...This page was left blank ...
Page 16: ...12 Figure 13 Pump ...
Page 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Page 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Page 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...