27
9.2.1.3
Flowrates
Press the
Flowrates
icon to check how many liters per minute of Air and CO
2
the system is consuming
(see Figure 39 a). Additionally, this panel allows you to select the flowrate of the gas mixture you want to send to
your incubator/device. Press the
Total
flowmeter (see Figure 39 b) to set the gas flowrate in the range 0.0÷1 L/min
by scrolling the
Total Flow
bar (1 in Figure 40) in the page that opens.
Figure 39. Flowrates View (a
–
b).
Figure 40. How to change the total flow rate.
Note
►
You can set the flowrate at 0.0 l/min, in this case the gas flowrate will be paused and output flow
will be interrupted (see
Error! Reference source not found.
). In this working mode three dots will appear i
n Homepage instead of CO
2
% current value and the indicator will turn yellow.
Note
►
When using the OKO-AIR-PUMP-BL as Air Source, the maximum flowrate value you can set is 0.4
l/min.