A-107 Color Measurement Instruments
ES3640e MFP Color Guide - 107
4. Plug the small connector on the adapter cable into the side of the
X-Rite DTP32 and plug the AC adapter into a wall outlet.
5. Turn on the computer.
6. Calibrate the DTP32 (see the following section).
7. Calibrate the Fiery using ColorWise Pro Tools and the DTP32 (see
Calibrating the DTP32
You will need the black-and-white X-Rite Auto-Cal Strip included
with the densitometer. Calibrating the densitometer does not
require ColorWise Pro Tools.
To calibrate the X-Rite DTP32
1. Connect the densitometer to the computer and supply power (see
).
2. From the Main Menu on the X-Rite DTP32 display, press the p1
key once to reach p2.
3. Press the cal key.
Calibrating motor speed
is displayed, followed by the words INSERT
CAL STRIP
.
AC adapter
Small connector