background image

SIDE

   

Before You Begin

PLEASE READ THIS BEFORE USING.

The following basic safety precautions should always be taken.

1.  Close supervision is necessary when the device is used by, on, or near children, 

handicapped persons or invalids.

2.  Use the device only for the intended use described in this User Guide.

3.  Do not use test strips and control solutions which are not supplied by the 

manufacturer.

4.  Do not use the device if it is not working properly, or if it has suffered any damage.

5.  Before using any product to test your blood glucose, read all instructions 

thoroughly and practice the test. Do all quality control checks as directed and 

consult with a diabetes healthcare professional.

6.  KEEP THIS USER GUIDE WITH YOU.

Intended Use

The system is intended for use outside the body (

in vitro

 diagnostic use only). 

It should be used only for testing blood glucose (blood sugar) and only with fresh 

capillary whole blood samples. The system is intended for use in the home and in 

clinical settings. It should not be used for the diagnosis of diabetes or for the testing of 

newborns.

     Getting To Know Your System

The OKmeter Direct Blood Glucose Monitoring System.

The OKmeter Direct system uses the latest technology to provide you with easy and 

comfortable testing. The system requires only a 0.7 μL of blood sample to complete 

the testing in only 6 seconds.
The OKmeter Direct system consists of

1. OKmeter Direct Blood Glucose Meter

2. OKmeter Direct Blood Glucose Test Strips

Principle of Measurement

Blood glucose is measured by an electrical current that is produced when a blood 

samples mixes with the reagent (special chemicals) of the test strip. The electrical 

current changes with the amount of glucose in the blood sample. The OKmeter Direct 

meter measures the strength of the electrical current, calculates your blood glucose 

level and then displays your result in either milligrams of glucose per deciliter (mg/dL) or 

millimoles of glucose per liter (mmol/L). 

Caution

1.  The user should not take any decision of medical relevance without first consulting 

his or her medical practitioner. 

2.  Call your doctor immediately if you experience symptoms that are not consistent with 

your blood glucose test results.

3.  High altitudes above than 3,402 meter (11,161 ft) may affect the test results. 

4.  Temperatures outside the range of 10°C to 40°C (50°F to 104°F) may affect the test 

results. Do not test beyond of temperature range.

IMPORTANT HEALTH-RELATED INFORMATION

1.  Apply only capillary whole blood sample to test your blood glucose. Applying other 

substances or plasma, serum will cause wrong results.

2.  Severe dehydration and excessive water loss may cause false low results. If 

you believe you are suffering from severe dehydration, consult your healthcare 

professional immediately.

3.  Test  results  below  60  mg/dL  (3.3  mmol/L)*1  indicates  low  blood  glucose 

(hypoglycemia). Test results greater than 240 mg/dL (13.3 mmol/L)*2 indicates high 

blood glucose (hyperglycemia).If your results are below 60 mg/dL (3.3 mmol/L) or 

above 240 mg/dL (13.3 mmol/L), repeat the test, and if the results are still below 60 

mg/dL (3.3 mmol/L) or above 240 mg/dL (13.3 mmol/L), consult your healthcare 

professional immediately.

4.  Inaccurate results may occur in severely hypotensive individuals or patients 

in shock. Inaccurate low results may occur for individuals experiencing a 

hyperglycemic-hyperosmolar state, with or without ketosis. Critically ill patients 

should not be tested with blood glucose meters.

5.  Abnormal red blood cell counts (hematocrit level below 20% or above 60%) may 

cause false results. Please consult your healthcare professional if you do not know 

your hematocrit level.

6.  Interference: Reducing substances occurring in the blood naturally (uric acid, 

bilirubin) or from therapeutic treatments (ascorbic acid, acetaminophen) will not 

significantly affect OKmeter Direct test results. However, elevated concentrations 

of these substances may affect test results. The compounds listed in the tables 

were found to have no affect at the concentration indicated.

Compounds

Highest concentrations 

tested at which no 

interference occured

Compounds

Highest concentrations tested 

at which no interference 

occured

Acetaminophen

 8.0 mg/dL (0.53 mmol/L)

Gentisic Acid

 5.0 mg/dL (0.32 mmol/L)

Ascorbic Acid 

 5.0 mg/dL (0.28 mmol/L)

Hydroxyurea

 3.0 mg/dL (0.39 mmol/L)

Aspirin

 60 mg/dL (3.33 mmol/L)

L-dopa

 10 mg/dL (0.51 mmol/L)

Bilirubin

 90 mg/dL (1.54 mmol/L)

Maltose

 900 mg/dL (26.3 mmol/L)

Cholesterol

 500 mg/dL (12.9 mmol/L)

Methyldopa

 3.0 mg/dL (0.13 mmol/L)

Creatinine

 5.0 mg/dL (0.44 mmol/L)

Tolbutamide

 400 mg/dL (14.8 mmol/L)

Dopamine

 2.0 mg/dL (0.11 mmol/L)

Triglycerides

 2,000 mg/dL (22.6 mmol/L)

Galactose

 900 mg/dL (50 mmol/L)

Uric Acid

 8.0 mg/dL (0.48 mmol/L)

REFERENCE:

* 1: Kahn, R. and Weir, G.: Joslinis Diabetes Mellitus, 13thed Philadelphia : Lea and Febiger (1994), 489.  
* 2: Krall, L.P. and Beaser, R. S.: Joslin Diabetes Manual. Philadelphia : Lea and Febiger(1989), 261-263.

1. TEST STRIP SYMBOL:

 Appears when the

    meter is turned on and waiting for inserting a

    test strip.

2. VOICE SYMBOL

3. BATTERY SYMBOL:

 Appears when battery is

    weak.

4. BLOOD DROP SYMBOL:

 Flashes when it is 

    ready to apply the blood sample.

5. MEASUREMENT UNIT:

 Appears with the test

    result in mg/dL or mmol/L.

6. Ketone?: 

Appears when result is higher than

    240 mg/dL (13.3 mmol/L).

7

.  : Appears when test result is within the range

    of 70 to 120 mg/dL (3.9 mmol/L to 6.7 mmol/L).

8.   AVERAGE:

 Appears in the memory mode while recalling 7/14/21/28 day test

      averages.

9.   MEMORY SYMBOL:

 Appears when in the memory mode.

10. CTL SYMBOL:

 Appears when doing a control sloution test.

11. TIME :

 The meter is programmed for a 12 hour period format.

12. DATE:

 Display month and date.

1

3.   AND LOW SIGN:

 Appears when test result is within the range of 20 to 70 mg/dL  

      (1.1 mmol/L to 3.9 mmol/L).

14.   AND HIGH SIGN:

 Appears when test result is higher than 120 mg/dL (6.7 mmol/L). 

15. THERMOMETER SYMBOL:

 Appears when ambient temperature is above or below 

      the acceptable range needed for testing. 

1.

A

1

. Getting To Know Your System

2. Prepare For Blood Sampling

3. Performing Blood Test
4. Understanding Your Test Result
5. Memory Recall
6. Control Solution Testing
7. System Specifications
8. Caring For Your Meter And Test Strip
9. Display Messages And Problem-Solving Guide

SIDE

  

A  :

   

SIDE

  

B  :

   

Test Strip

ABSORBENT 

CHANNEL:

Apply a drop of 

blood here.

CONTACT BARS:

Insert this end into strip slot.

TEST STRIP HANDLE :

Hold this part to insert the test strip 

into the meter slot.

CONFIRMATION WINDOW:

This is where we confirm if enough blood is applied for the 

test. Ensure that it is fully covered with blood sample.

How to replace the battery

 

1. Make sure the meter is turned off. Let the front of the meter 

rest in the palm of your hand. Slide battery cover open.

S

2. Remove the old batteries and replace with two (2) new 1.5V 

AAA size alkaline batteries.

 

3. Close battery cover.  

NOTE: 

1. Replacing the battery does not affect the test result stored in memory. However the 

time and date may need to re-set.

2. As with all small objects, the battery should be kept away from small children as a 

safety precaution. If the battery is swallowed, seek medical assistance immediately.

3. Batteries might leak chemicals if not used for a long time. Remove the batteries if 

you are not going to use the device for extended (i.e., 3 months or more ).

4.  Please discard the product or the batteries properly according to the regulations 

of your country.

For 

In vitro diagnostic use

Temperature limitation / Store at

 

Please consult instructions 

for use

 

Use by /Expiry date

 

Do not reuse

 

Manufacturer

 

Lot number

EU representative

 

Caution, consult 

accompanying document

 

This product fulfils the requirements of 

Directive 98/79/EC 

In vitro diagnostic 

medical device.

Performance Characteristics:

ACCURACY: More than ninety-five percent (95 %) of the individual glucose results fell within ± 0.83 mmol/L (15 mg/dL) of the 

results of the glucose analyzer YSI 2300 measurement procedure at glucose concentrations < 4.2 mmol/L (< 75 mg/dL) and 

within ± 20 % at glucose concentrations

 

 

4.2 mmol/L (

 

75 mg/dL). 

PRECISION: CVs (%) of intermediate precision and repeatability were less than 5%.

1

3

2

4
5
6

7

8

11

12

13

14

15

10

9

Code

REPLACING THE BATTERY

The OKmeter Direct meter comes with TWO 1.5V AAA alkaline batteries.
Battery life will vary depending on usage, so always keep spare batteries on hand.
The battery should last about 1000 tests or 12 months when testing 3 times a day.
When the batteries symbol appears on the meter display, battery is getting low. You will 

still be able to test with low battery, but you should replace them as soon as possible. 
When battery symbol and E-b shows up in the meter display, the meter will no longer 

give results and you must replace the batteries immediately. Please always have spare 

batteries with you to ensure that you can replace the battery anytime.

SETTING TIME AND DATE

Please install battery first and set correct time and date before you begin to test.

SET

STEP 1. Enter Setting Mode

Start with the meter off. Open the battery compartment, and press SET 

BUTTON. The meter will turn on and enter the setting mode. 

STEP 2. Set the Year

The year appears with the number flashing. Press and release the M 

BUTTON to obtain the desired year. Press SET BUTTON to confirm 

and move to next MONTH setting.

STEP 3. Set the Month

The month appears with the number flashing. Press and release the M 

BUTTON to obtain the desired month. Press SET BUTTON to confirm 

and move to next DATE setting.

STEP 4. Set the Date

The date appears with the number flashing. Press and release the M 

BUTTON to obtain the desired date. Press SET BUTTON to confirm 

and move to next HOUR setting.

STEP 5. Set the Hour

The hour appears with the number flashing. Press and release the M 

BUTTON to obtain the desired hour. Press SET BUTTON to confirm 

and move to next minutes setting.

STEP 6. Set the Minutes

The minute appears with the number flashing. Press and release the M 

BUTTON to obtain the desired minute. Press SET BUTTON to confirm 

and the meter will enter to next unit setting.

STEP 7. Set the Unit of Measurement

The existing unit (for example mg/dL) will appear and flash. You can 

choose to reset measurement unit. If you do not want to change unit, 

press SET BUTTON to skip this step. If you want to change unit, press 

and hold the M BUTTON for 3 seconds to change the unit to mmol/L. 

Press SET BUTTON to confirm and the meter will enter to next setting.

STEP 8. Set the Unit of Temperature

The existing unit (for example °C) will appear and flashing. Press the M 

BUTTON to change the unit to °F. Press SET BUTTON to confirm and 

the meter will enter to next delete all memory setting.

STEP 9. Delete Memory

When the dEL symbol and the flashing memory symbol appear on the 

display, you can choose to clear the memory. If you do not want to clear 

the memory, press the SET BUTTON again to skip this step. If you want 

to clear ALL memory, press and hold M BUTTON for 3 seconds. The “--

-“ image will appear on the LCD screen to indicate that all memory has 

been deleted. 

STEP 10. Complete Setting

After delete memory, the meter will display “OFF” before shut down. 

The meter setting is now completed.

IMPORTANT :

 

1.The time, date and unit of measurement can ONLY be changed in the setting mode. Therefore, 

when you perform a blood glucose testing, it is not possible to change those parameters.

2.Your meter displays test results in milligram per deciliter (mg/dL) or millimoles of glucose per 

deciliter. Use of the wrong unit of measure may cause you to misinterpret your blood glucose 

level, and may lead to incorrect treatment. Please always consult with your healthcare 

professionals before you reset the unit of measure.

3.While the meter is in the setting mode, if no button is pressed for 1 minute, the meter will turn off 

automatically. 

     Prepare For Blood Sampling

Adjustable Lancing Device 

Your lancing device and lancets are used for obtaining

capillary blood samples from the puncture site.

Lancet

Setting your Lancing Device

 

 

1. Screw off the cap of lancing device. Insert a 

lancet into the lancet holder and push down until 

it is fully seated.

2. Twist off the protective cap until it separates from 

the lancet.

3.  Replace the lancing device cap and set the  

     puncture depth to the desired number.

      

To select the best depth:

     

 For delicate skin

     

 For normal skin

     

 For thick or callused skin

 

 

4. Pull back the Sliding Barrel until it makes a click, 

and then release. If it does not click, the device 

may have been cocked when the lancet was 

inserted.

Protective cap

2

.

 Important Lancing Device and Lancets Information

1.  Lancet is for single use only.

2.  Keep lancing device and lancets clean.

3.  Use caution when removing the used lancet from the device and when disposing the 

used lancet.

IMPORTANT : 

The meter and lancing device are for single patient use.

                     Do NOT share them with anyone including other family members !

                     Do NOT use on multiple patients !

User Guide

Blood Glucose Monitoring System

Blood Glucose Monitoring System

New

10-62-6219-0003   V1-AUG13

OK Biotech Co., Ltd.  

          

4F-1, No. 83, Sec. 2, Gongdao 5th Rd., 

Hsinchu City 30070, Taiwan

TEL: +886-3-5160258

FAX: +886-3-5160028

www.okbiotech.com

E-mail: [email protected] 

MDSS GmbH

Schiffgraben 41, D-30175 

Hannover, Germany

Meter

S

OK1NJ005001

Model Name: OK-1N  Made in Taiwan

Battery : 1.5 V AAA x 2  

OK Biotech Co., Ltd.

www.okbiotech.com

           For self-testing.

10-62-6263-0103  V1-AUG13

DISPLAY :

The large, easy 

to read display 

shows test results, 

messages, test 

results stored in 

memory.

STRIP SLOT :

Insert the test strip here. 

The meter will turn on 

automatically.

SET BUTTON :

Located in the battery compartment; press button 

to enter meter setting. Refer to “SETTING YOUR 

METER” section of this use guide for details.

M BUTTON :

Main button, press button to turn the 

meter on or perform other functions 

described in this manua

l.

DATA PORT :

Allows you to transfer the information stored in the 

meter to a computer to view, analyze and print. 
Please contact us if you need detailed information.

METER LABEL :

Each meter has its unique 

number on it. Do not alter 

or tear the label off.

BATTERY 

COMPARTMENT :

Holds TWO 1.5V 

AAA alkaline 

batteries.

Arrow

Lancet Holder

Depth Selectors

(Offer 6 levels of skin penetration)

Release Button

Lancet Ejector

Sliding Barrel

Lancing Device Cap

STRIP EJECTOR :

Slide the ejector forward 

to discard the test strip.

Reviews: