background image

direct sunlight. 

4.

After 2 hours check the instrument reading; it should be 12% ± 2%. 

          
         If readjustment is necessary: 
         Model 509-1/2:   Remove 4 screws on back of instrument; gently open hous- 
                                    ing.  See Figure 1 for pot location. 
         Model 509-3/4:   Remove screw from probe housing and gently pull out the   
                                    PC board.  See Figure 2 for the location of the potentiome-  
                                    ters.  To readjust the pots without removing the probe hous-
                                    ing you can drill through the two prepared holes, situated in 
                                    the middle of the housing. 
 
5.

Adjust the zero potentiometer for a reading of 12% RH. 

6.

Replace the 12% RH bottle with the 90% bottle.  Close the 12% bottle. 

7.

Wait for 2 hours and check instrument reading; it should be 90%, ± 2%. 

8.

If necessary, readjust the span potentiometer for a 90% RH reading. 

9.

Repeat Steps 5 through 8 until both readings are correct; when that happens 
the meter will meet its original accuracy specification.  

Close the bottles im-

mediately after use and wash your hands; the bottles can contain poisonous 
solutions.

  Reassemble the housing. 

 

 
 
Calibration Standards 

 
Each unit is calibrated against standards traceable to the National Institute of 
Standards and Technology. 
 
 

Calibration certificates 

 
In addition to the normal calibration procedure, each unit can be supplied with its 
own traceable calibration certificate. 
 
 

Calibration interval time 

 
Under normal ambient conditions (0 to 50°C, 0 to 70% RH) and for an accuracy of  
± 2% RH, we recommend an annual calibration.  For an accuracy ±5% RH we 
recommend calibration every five years.  For environments with airborne chemi-
cals or for high humidity and high temperature conditions we recommend more 
frequent calibration. 
 
 

Limited Warranty 

 
This product is warranted by Ohmic Instruments Company to be free of defects in 
material and workmanship for one year after delivery.  A product found to be de-
fective for these reasons within this period will be repaired or replaced free of 
charge by Ohmic.  We give no other warranties.  Ohmic Instruments Co. shall not 
be liable for any damages or losses, whether direct or indirect.  The warranty can-
not be transferred or assigned to third parties. 

508 August Street    Easton, MD  21601 

www.ohmicinstruments.com 

Voice  (410) 820-5111 

Fax  (410) 822-9633 

Instruction Manual for Relative Humidity 

& Temperature Meter    DM-509 

Introduction 

 

The DM-509 series of digital handheld meters are designed to indicate the relative 
humidity or temperature of the ambient air.  Featuring auto power-off and a large 
3½ digit LCD display, the meter is standard equipped with a fast reaponse capaci-
tive sensor with a working range of 10 to 95% RH.  Accessories like calibration 
bottles or carrying cases are available from Ohmic. 

Description 

 

 

Digital meter with fixed probe, temperature in °C 

DM-509-M-01 

—— 

Digital meter with fixed probe, temperature in °F 

DM-509-M-02 

—— 

Digital meter with separate probe, temperature in °C  DM-509-M-03  DM-509-T-03 

Digital meter with separate probe, temperature in °F  DM-509-M-04  DM-509-T-04 

Model 