background image

2-4

6600-0371-000

10/27/03

2/Setup and Checkout Procedures

2-4

2.3 Control Unit Checkout Procedure

WARNING:

w

Do not perform the Checkout Procedures (Mechanical and Control Unit)
while a patient occupies the warmer.

w

Complete the “Checkout Procedures” section of this manual before
putting the unit into operation. If the warmer fails any portion of the
checkout procedures it must be removed from use and repaired.

A. Control Unit Check

1.

Connect the warmer power cord to an appropriate power source. Refer to the rating
plate on the Warmer for the proper voltage needed. Switch the power On and verify
the following on the Control Panel (Figure 2-2):

a. The alternating two tone audible alarm sounds and all displays and indicators are

lit for approximately two seconds.

Note: During this time the controller also performs self check functions. If the
controller detects a failure, the alarm stays on and service is required.

Note: All alarms except system failure or power failure are preceded by a 30
second intermittent operator prompt tone.

b. The display flashes "PrE HEAT" until the mode, increase/decrease or Apgar

touch switches are pressed. Press the decrease touch switch.

c. The manual mode indicator is lit.

d. Operator prompt tones sound and the % power display flashes.

2.

Adjust the heat output with the increase (

M

) and decrease (

?

) touch switches to

attain the high and low limits as indicated by the % power display.

3.

Connect the skin temperature probe to the warmer.

4.

Press the mode touch switch to place the warmer in the servo mode and verify the
following:

a. The servo mode indicator is lit.

b. An operator prompt tone sounds and the control temperature display flashes 36.5°C.

5.

Press the increase (

M

) touch switch and verify that the maximum servo control

temperature attainable is 37.5°C.

Note: A patient temperature alarm occurs if the difference between the patient
temperature and the control temperature is greater than 1°C (the difference can be
adjusted to 0.5°C by a qualified service person).

Note: An alternating two tone alarm, a flashing overhead alarm light and a flashing
patient temperature display may occur here if the skin temperature probe is below
30°C. Warm the probe with your fingers or silence the alarm.

6.

Press the decrease touch switch and verify that the minimum servo control tempera-
ture attainable is 35.0°C.

7.

Disconnect the skin temperature probe. Verify the following:

a. The probe failure indicator light is lit.

b. There is an alternating two tone alarm.

c. The overhead alarm light is flashing.

6600−0371−000−DWG

100

MANUAL−DOC, User Reference Manual, OM Panda, English−default

Reproduced from the electronic master in MATRIX

Summary of Contents for Panda

Page 1: ...mer OperationandMaintenanceManual 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX Approved 2009 6 4 16 13 EET DST Barbar...

Page 2: ...to the nearest Ohmeda Medical Regional Service Center This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Ohmeda Medical and by Ohme...

Page 3: ...CheckoutProcedure 2 4 A Control Unit Check 2 4 B ElapsedTimerCheck 2 5 C ObservationLightCheck 2 5 D Raise and Lower Bed Switch Check 2 5 E Interlock Switch Check 2 6 F Power Failure Memory and Batter...

Page 4: ...y 5 5 5 5 Battery Test 5 5 5 6 Maintenance Schedule 5 6 A Operator Maintenance 5 6 B Service Maintenance 5 6 6 Illustrated Parts Appendix Temperature Conversion Chart A 1 Servo Mode Algorithm A 2 Infa...

Page 5: ...is used when the possibility of injury to the patient or the operator exists Symbols Attention Consult the manual for more information Type B equipment FunctionalEarthTerminal ProtectionEarthTerminal...

Page 6: ...iv 6600 0371 000 10 27 03 Table of Contents iv Notes 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX...

Page 7: ...rocedures Mechanical and Control Unit while a patient occupies the Warmer Complete the Checkout Procedures section of this manual before putting the unit into operation If the warmer fails any portion...

Page 8: ...vi 6600 0371 000 10 27 03 Table of Contents vi Notes 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX...

Page 9: ...accessory items 1 3 Heater Assembly Thewarmerisdesignedtowarminfantsonabedsurfacelocatedbeneaththeheaterassembly WARNING w Bed to heater spacing less than the recommended 64 cm will result in incor r...

Page 10: ...le bed platformallowsTrendelenburgandreverseTrendelenburgpositioning Thehydraulic system for the tiltable bed provides a smooth dampened motion to avoid disturbing the patient 1 6 Bassinet Two bassine...

Page 11: ...rocedure WARNING w Before using the Ohmeda Panda Baby Warmer read this entire manual Attempting to use this device without a thorough understanding of its operation may result in patient or user injur...

Page 12: ...back to the normal position Check for a smooth rotation C Mechanical Checks 1 Check the operation of the bed sides The bed sides should operate smoothly WARNING w Regularly inspect the bed side panel...

Page 13: ...ator Alarm Lights Alarm Silence Mode Switch Mode Indicators Bed Controls Bed controls present on elevating bed models only CI 14 096 Figure 2 1 Wall mount pin insertion 1 Verify that the warmer is rig...

Page 14: ...prompt tones sound and the power display flashes 2 Adjust the heat output with the increase M and decrease touch switches to attain the high and low limits as indicated by the power display 3 Connect...

Page 15: ...ch switch to activate the elapsed timer Verify that the timer starts operating 2 Press the Apgar tones On Off touch switch for the Apgar tones Verify that the indicator light for the Apgar tones is ex...

Page 16: ...10 minutes test or partially charged to pass the two minute test If the battery is defective replace it Refer to the service manual There is no maintenance required for the battery The battery has a...

Page 17: ...erature sensed at the skin tempera ture probe tip The skin temperature probe must be properly connected to the warmer and the infant for an accurate patient temperature measurement The range of temper...

Page 18: ...st take the responsibility for making heater adjustments or environmental changes in response to the following conditions patient size and weight bed to heater head height differences changes in the e...

Page 19: ...alarm tone is sounded at one minute and every five minute interval after the elapsed timer is started LightOn Off This switch activates the observation light located in the heater assembly WARNING w...

Page 20: ...owerFailureAlarm Probe Failure Alarm The probe failure alarm is only active in the servo mode of operation The alarm activates when the skin temperature probe fails electrically due to an open or shor...

Reviews: